• Compact, cost-effective endoscope reprocessing featuring simple, push-button operation • Fast cycle time for quick endoscope turnaround • Individual endoscope channel connections ensure disinfectant and rinse water contact with all channels • Pre-programmed cycles for reprocessing cycle repeatability and consistency • CER-1 model disinfects one endoscope per cycle • CER-2 model disinfects up to two endoscopes per cycle • Gasket lid and sealed reservoir contain disinfectant vapors • Alcohol flush and air drying of endoscopes • Audible and visual alarms for reprocessing cycle deficiency • Assures patient and operator safety • Self-disinfects internal fluid lines during every cycle • Bacteria retentive two-stage water filtration system • Compatible with all current models of endoscopes, including those with elevator wire channels • For use with most FDA-cleared reusable high-level disinfectants • Vapor management system, disinfectant transfer pump, neutralization system and mobile cart optional
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Under/over-dosed disinfectant; cycle fails self-test.
Root cause: Worn peristaltic pump or clogged dosing line.
Symptoms: Cycle aborts on leak or connectivity test.
Root cause: Worn connector fitting or a blocked scope channel.
Symptoms: Failed rinse-water quality alarm.
Root cause: Saturated filter or UV lamp failure.
Symptoms: Won't start cycle, door won't seal.
Root cause: Worn door gasket or interlock switch.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable; replace per cycle count.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Water filter cartridge (bacterial/particulate) Consumable; replace per cycle count. | Frequent wear item | $$ | Easy |
| UV lamp (where equipped) | Frequent wear item | $$ | Fairly easy |
| Door gasket / seal | Periodic replacement | $ | Fairly easy |
| Chemical dosing pump | Periodic replacement | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.