Olympus · Model OER Pro
Quick Reprocessing Time. Simultaneous reprocessing of two flexible scopes in about 29 minutes, or one scope at a time, including Olympus ultrasonic scopes and other specialty models. Compact and Smart Design. Simple configuration in a compact 18"-wide designhalf the size of earlier reprocessorshelps ease operation and installation while minimizing space requirements. Also equipped with casters for increased mobility and accessibility. Modified Manual Cleaning. Easier and faster cleaning process, which eliminates 7 of the 11 manual cleaning steps, including the most labor-intensive and variable part of the process: manual flushing of the endoscope channels with detergent, water, and air. Automated Scope ID System. Equipped with a simple-to-use Radio Frequency Identification (RFID) management system, which provides automated traceability of the scope serial and model numbers, operator, and time of reprocessing to eliminate cumbersome manual input from a keypad or barcode. Data Management Simplify data input, streamline accurate record keeping, and reduce the need for paper receipt searches with two dedicated data management solutions that can be scaled to meet any busy reprocessing center's budgetary needs. OER-Pro data management solutions are focused on mitigating risk and achieving superior reprocessing compliance. Contact your local Olympus sales representative for more information on the Olympus Reprocessing Management Module and Data Capture Unit.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Cycle stops with a fault code, scope not high-level disinfected, must be re-run.
Root cause: Water fill / drain fault, a leak-test failure, a channel-flow / pressure fault, or a chemistry-delivery fault.
Symptoms: Low-flow or occlusion alarm on a scope channel, incomplete flushing.
Root cause: Wrong or worn hook-up connector, a blocked channel, or a flow-sensor fault.
Symptoms: The reprocessor's automated leak test fails.
Root cause: A genuine scope leak, or a worn leak-test connector / pump.
Symptoms: Low-concentration alarm, MEC test fails, dosing alarm.
Root cause: Empty or expired disinfectant, a dosing-pump fault, or a concentration-monitor fault.
Symptoms: Water-quality alarm, poor rinse, filter-change overdue.
Root cause: Loaded pre-filter / final micron filter, or an RO / water-supply issue.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Water filters (pre-filter & final micron) | Frequent wear item | $$ | Easy |
| Scope hook-up / channel connectors | Frequent wear item | $$ | Easy |
| Dosing / circulation pump | Periodic replacement | $$$ | Difficult |
| Leak-test pump / connector | Periodic replacement | $$ | Moderate |
| Door / chamber seal | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.