Medical Measurement Systems · Model Solar Blue
Small - Solar Blue requires very little space. It can be used as a table-top system but it can also be mounted on an investigation chair, a robust purpose-designed pole, or a wall. The result is a 100% set-up versatility and maximum convenience for physician, nurse, and patient. Wireless - Solar Blue is the only 100% wireless Bluetooth® urodynamic system currently available. Wireless between Solar Blue base unit and: flowmeter, UPP puller, remote control and computer. The Solar Blue base unit runs on rechargeable batteries. Result: no more cables, easy movement, more space and simple cleaning of individual modules. Easy - Solar Blue software is highly intuitive and logical, requiring minimal training. Includes an efficient patient database for easy storage and retrieval of investigation data. A purpose-designed remote control enables the investigator to stay close to the patient. Using just 4 buttons, the complete urodynamic examination can be carried out by 1 person. This shortens examination time, improves quality of the urodynamic investigation and provides greater patient privacy.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Vesical, abdominal or detrusor pressure offset, will not zero, implausible traces.
Root cause: Transducer not open to atmosphere / not at the correct reference during zero, a damaged transducer / dome, or an amplifier fault.
Symptoms: Filling volume inaccurate, pump will not run at the set rate, occlusion alarm.
Root cause: Peristaltic pump rollers / tubing, or a pump-drive fault.
Symptoms: Flow trace noisy, zero drift, weight / rotating-disc flowmeter inaccurate.
Root cause: Flowmeter load cell / disc, or a calibration fault.
Symptoms: Noisy or absent pelvic-floor EMG.
Root cause: Electrode contact / cable or the EMG amplifier.
Symptoms: Software hangs, study will not save or export.
Root cause: Workstation PC or software fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Pressure transducers / domes | Periodic replacement | $$ | Fairly easy |
| Peristaltic pump rollers / tubing | Frequent wear item | $$ | Moderate |
| Uroflow flowmeter | Major service part | $$$ | Moderate |
| EMG / patient cable set | Frequent wear item | $$ | Easy |
| Workstation PC | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.