MAMalvestio · Model TALENT
In case of emergency, the backrest can be brought down manually through a (CPR manoeuvre) quick release lever available on both sides of the bed; the levers are easily identifiable in all conditions of use. Mattress platform with thermoformed ABS covers, easily removable and sanitised. The mattress platform is equipped with an integrated bed lengthener, stratified laminate (HPL), manually extractable by 30 cm. The bed platform increases from 177 cm (in closed position) to 207 cm. Easily removable headboard and footboard, with a lever lock/unlock device. Made of polyethylene with stainless steel connection tubes. Prepared for the installation of a monitor shelf. Designed to hook the monitor shelf. The base, obstacle free, is covered by a plastic casing. Twin wheels, 150 mm diameter, with antistatic system, which can be locked simultaneously, complete the base. By acting on the pedals, located at the foot end, it is possible to unlock/lock the wheels or switch to directional mode. An acoustic warning will be activated in the event of a power connection and unbraked wheels. The bed is equipped with 4-section side rails, with integrated controls, with dampened movement and adouble-action lock/unlock system. The panels are made of polyethylene and their shape is designed to guarantee a containment height of 45 cm above the bed mattress platform.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Backrest, knee, height, Trendelenburg or the boost / turn assist will not move from the siderail or caregiver controls.
Root cause: Failed linear actuator, a control-board fault, a control-panel fault, or a pinched / damaged control cable.
Symptoms: Bed will not brake or release, drifts, central brake pedal ineffective.
Root cause: Worn central-brake linkage or castor.
Symptoms: Weight reads implausibly or will not zero, bed-exit alarm false-triggers or fails.
Root cause: Load-cell drift or damage, or a scale-control fault.
Symptoms: Functions and alarms do not work during a power interruption, battery-fault indicator.
Root cause: Aged back-up battery or a charger fault.
Symptoms: Siderail will not latch, rattles, or the CPR release is slow / stiff.
Root cause: Worn siderail latch or the CPR mechanism.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Recalibrate the scale after replacement.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Linear actuator (backrest / knee / height) | Major service part | $$$ | Difficult |
| Central-brake / castor assembly | Major service part | $$ | Difficult |
| Load cell / scale sensor set Recalibrate the scale after replacement. | Major service part | $$$ | Difficult |
| Back-up battery | Periodic replacement | $$ | Fairly easy |
| Siderail control panel / control board | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio