Kuraray Medical · Model KPS-8800Ce
Applicable to eleven treatment systems. Self-washing and self-recovery. Many safety control functions. Automatic airtight test for blood circuit. Double check on blood recovery lines. Detection of the leakage from Rolling tube.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Blood, plasma or replacement-fluid pump flow error, occlusion or pressure alarm, inaccurate balance.
Root cause: Worn pump rollers / race, a pump-door / occlusion-sensor fault, or a motor / encoder fault.
Symptoms: Inlet, return, plasma or TMP pressure reads wrong; frequent pressure alarms.
Root cause: A pressure transducer drifted or a protector wetted.
Symptoms: Air-in-line or blood-leak alarm, pump stops, return clamp closes.
Root cause: Air entrained at a connection, a dirty detector optic, or a detector fault.
Symptoms: Fluid-balance alarm, implausible plasma / replacement volume.
Root cause: A weighing scale / load cell out of calibration or mechanically bound.
Symptoms: Return fluid too cool or warm, temperature alarm.
Root cause: Heater element or temperature sensor.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Recalibrate fluid balance after replacement.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Peristaltic pump roller / race kit | Periodic replacement | $$$ | Difficult |
| Pressure-transducer / protector set | Periodic replacement | $$ | Moderate |
| Blood-leak / air detector | Major service part | $$$ | Difficult |
| Weighing scale / load cell Recalibrate fluid balance after replacement. | Major service part | $$$ | Difficult |
| Pre-connected plasma-separator tubing set | Consumable | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio