
International Biomedical · Model Model 185A+
Double- wall hood with full 360° visibility. Head access door. Front access door with two hand ports. Accessory ports for IV/ respiratory tubing. Mattress. High-intensity exam light. Skin temperature probe. Humidification reservoir. IV pole. Infant restraint straps. 2 D or 2 E cylinder mounts.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Chamber / patient will not reach or hold the set temperature, temperature alarm, self-test fault; "Check" test shows a failed LED / lamp / buzzer.
Root cause: Heater element / infrared rod, the air- or skin-temperature sensor, the fan (incubator), or a control-board fault.
Symptoms: Servo mode does not regulate to the skin probe, deviation alarm, probe-fault alarm.
Root cause: Damaged skin temperature probe or its connector, or the servo-control circuit.
Symptoms: No or quiet audible alarm, central alarm lamp does not light on the "Check" test.
Root cause: Buzzer / speaker, the central alarm lamp, or the alarm board - a safety-critical failure.
Symptoms: Mattress heating uneven or not warming, mattress-temperature alarm.
Root cause: Mattress heating element or its sensor.
Symptoms: Set humidity not reached, uneven chamber temperature, airflow alarm.
Root cause: Humidifier reservoir / heater, or a worn circulation fan / blocked filter.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Air / skin temperature sensors | Frequent wear item | $$ | Fairly easy |
| Heater element / infrared rod | Major service part | $$ | Difficult |
| Circulation fan (incubator) | Major service part | $$ | Moderate |
| Heated mattress / gel pad (warmer) | Periodic replacement | $$ | Fairly easy |
| Buzzer / central alarm lamp | Major service part | $ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio