Imex Medical Systems · Model ImexLab 9100
• Better Patient Care - Test patients over 65, hypertensives, diabetics, smokers and victims of stroke, TIAs or heart attack when PAD symptoms appear and prevent more serious complications or surgeries. • Full Range of Modalities - The systems performs a wide range of noninvasive test including arterial pressure studies, bilateral photoplethysmography (PPG) and bilateral pneumoplethysmography for waveform analysis. • Disk Storage: - Store test results on disk until you are ready to print. The system generates full-page, color reports in seconds with CPT and ICD-9 codes if desired.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Weak, absent, or noisy Doppler signal.
Root cause: Probe crystal or cable wear.
Symptoms: Inaccurate or failed pressure readings.
Root cause: Cracked cuff bladder or tubing/connector leak.
Symptoms: Distorted or flat plethysmography waveform.
Root cause: Pneumatic transducer calibration drift.
Symptoms: Incorrect ABI/TBI calculation, or frozen screen.
Root cause: Application software or workstation fault.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
High-turnover; keep spares.
Consumable; multiple sizes.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Doppler probe High-turnover; keep spares. | Periodic replacement | $$$ | Fairly easy |
| Blood pressure cuff set Consumable; multiple sizes. | Periodic replacement | $ | Easy |
| Pneumatic tubing / connector set | Periodic replacement | $$ | Fairly easy |
| Workstation / PC | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →