Image Works · Model Panoura 18 S
The Panoura 18S is the most practical advanced 2D and 3D dental imaging system. Panoura 18S 3D is designed to be an ideal model for 3D panoramic systems. Slim, compact and highly functional. Adult Pan, Child Pan, Bite Wing, FMX, TMJ, Ceph, 3D imaging for every dental practice. The Panoura 18S packs a lot of technology into a small footprint, utilizing a high-end CMOS sensor that can be transferred to an optional cephalometric arm. With a user-adjustable focal trough, errant images can be brought back into focus after acquisition. At various speeds, the Panoura 18S can take a quick pan of 6 seconds, up to a 16 second pan for maximum resolution. The Panoura 18S is bundled into an aesthetically pleasing package that enhances the appearance of any office
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Incomplete or jerky rotation arc; scan aborts.
Root cause: Worn rotation motor or drive belt/gear.
Symptoms: Banding or uneven density across the scan arc.
Root cause: CCD/CMOS sensor or slit calibration drift.
Symptoms: Repeated rejects on start, positioning-light misalignment.
Root cause: Bite-block/laser positioning sensor drift.
Symptoms: Ceph images misaligned, or arm won't lock in position.
Root cause: Telescoping arm joint or cable wear.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Only on ceph-equipped units.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Cephalometric arm assembly Only on ceph-equipped units. | Major service part | $$$ | Difficult |
| Rotation drive motor / belt | Major service part | $$$ | Difficult |
| Panoramic sensor / detector | Major service part | $$$$ | Expert-level |
| Positioning laser / light module | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →