Hitachi Medical Systems · Model HI Vision 5500
Ease of Use Tri-color keyboard lighting for intuitive operation One Touch Access to Commonly Used Functions User Defined Programmable Keys Durable, Lightweight, and Highly Mobile Advanced Performance Quadra Beam Processing for Fast Frame Rates 12 bit Gigasampling A/D for Precise Signal Reproduction Windows XP based for reliability, stability, and familiarity HI VUE circuitry for streamlined operation and signal flow 3 Modes of dynamic Tissue Harmonic Imaging (dTHI) Directional Color Flow Angiography Real-time Doppler Measurements Standard Features Patient and Image Database Management System Enhanced Cine and M-mode recording with Cine-max memory 3RD Generation Color Artifact Suppression Powerful Built-in Measurement and Equation Editor Customizable Presets and Preset-Operation Functions Optional Imaging Enhancements WideView Panoramic Imaging Option 3D Imaging Options (built-in or stand-alone) Single Omni-directional M-Mode Display Option Steerable CW Doppler Option ECG Biophysical Display Option Pentax EUS and Fujinon Mini-probe Options.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Noisy, low-contrast or non-uniform image, missing sector, drop-out that follows the system not the probe.
Root cause: Front-end / beamformer board fault, a probe-port (connector) board fault, or a system-board fault.
Symptoms: A probe works on one port but not another, or is not detected at all.
Root cause: Worn probe-port connector or a scan-connector board fault.
Symptoms: Dead system, hangs at boot, or powers off under load.
Root cause: Power supply / DC converter, the system battery, or a system-board fault.
Symptoms: Unresponsive TGC sliders or keys, distorted or blank display.
Root cause: Control-panel board, touchscreen, or the monitor.
Symptoms: Thermal-shutdown warning, fan roar, hot exhaust.
Root cause: Failed or dust-clogged cooling fan or blocked air path.
Symptoms: Studies will not send to PACS, images will not archive, worklist empty.
Root cause: Network configuration or a storage-drive fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Front-end / beamformer board | Major service part | $$$$ | Difficult |
| Probe-port / scan-connector board | Major service part | $$$ | Difficult |
| Power supply / DC converter | Major service part | $$$ | Difficult |
| Control-panel / touchscreen assembly | Major service part | $$$ | Moderate |
| Cooling fan | Major service part | $ | Moderate |
| System hard drive | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.