Hitachi Medical Systems · Model EUB 8500
[u]Powerful Processor[/u] The outstandingly powerful processor of the HITACH EUB-8500 permits simultaneous processing of highly complex functions at high speed, allowing anatomical and physiological details to be displayed with unprecedented accuracy. [u]High Performance Transducers[/u] The HITACHI-EUB-8500 offers in excess of 30 high performance transducers addressing a full range of diagnostic applications, either for mainstream or specialized examinations. Ergonomics Enhanced ergonomics ensure user comfort and intuitive access to all functions within the system. With its extremely compact and light weight design the HITACHI EUB-8500 can easily be maneuvered in even the most difficult situations. Enhanced Productivity 'The Workflow-Concept'. Whilst physicians and patients are satisfied with shorter examination times together with a high reliability of diagnostic findings, the Management and Administration Departments appreciate the increased efficiency of the workflow pattern - all enabled by the HITACHI-EUB-8500. Patient data can be retrieved from the hospital´s network by simply pressing a button. The application specific set-up of the system in Real Time, its intuitive operation, the high-speed switching between imaging modes, quick archiving of images, easy access to stored images as well as the possibility to store video clips ensure a high patient throughput which leads to greater efficiency of the clinical workflow. Enhanced productivity and economic efficiency in conjunction with a user-friendly operating system based on the latest ergonomic principles.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Noisy, low-contrast or non-uniform image, missing sector, drop-out that follows the system not the probe.
Root cause: Front-end / beamformer board fault, a probe-port (connector) board fault, or a system-board fault.
Symptoms: A probe works on one port but not another, or is not detected at all.
Root cause: Worn probe-port connector or a scan-connector board fault.
Symptoms: Dead system, hangs at boot, or powers off under load.
Root cause: Power supply / DC converter, the system battery, or a system-board fault.
Symptoms: Unresponsive TGC sliders or keys, distorted or blank display.
Root cause: Control-panel board, touchscreen, or the monitor.
Symptoms: Thermal-shutdown warning, fan roar, hot exhaust.
Root cause: Failed or dust-clogged cooling fan or blocked air path.
Symptoms: Studies will not send to PACS, images will not archive, worklist empty.
Root cause: Network configuration or a storage-drive fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Front-end / beamformer board | Major service part | $$$$ | Difficult |
| Probe-port / scan-connector board | Major service part | $$$ | Difficult |
| Power supply / DC converter | Major service part | $$$ | Difficult |
| Control-panel / touchscreen assembly | Major service part | $$$ | Moderate |
| Cooling fan | Major service part | $ | Moderate |
| System hard drive | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.