Hillrom · Model UniFlex Headwalls
Uniflex® Headwalls Easy Installation: Delivered pre-plumbed and pre-wired to permit single point connections for each service type. Flexible Design: The headwall comes in variable lengths and a wide choice of laminate colors.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Hiss at an outlet, low-pressure alarm on bedside equipment.
Root cause: Worn outlet check valve/seal or a loose union in the pre-plumbed manifold.
Symptoms: Dead or intermittent outlet, breaker trips.
Root cause: Loose pre-wired terminal or worn receptacle.
Symptoms: Panels bind, gaps at joints between headwall lengths.
Root cause: Frame racking or worn panel hardware.
Symptoms: Nurse-call not registering, dead data jack.
Root cause: Unseated punch-down or damaged cable.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Gas outlet check-valve / seal kit (per gas) | Major service part | $$ | Moderate |
| Hospital-grade receptacle | Major service part | $ | Fairly easy |
| Access-panel hardware kit | Major service part | $ | Moderate |
| Nurse-call / data jack module | Major service part | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.