Hillrom · Model Contour Headwall System
Contour® Headwall from Hill-Rom: Clean, ergonomic design, clinically flexible and visually compelling Hill-Rom® Contour® Headwall Systems The new. When was the last time you could say that about a headwall? Clean, Ergonomic Design The gases, electrical outlets, data and nurse call are all in the ergonomic strike zone, so there is no need to bend or reach. The vertical equipment management tracks help manage and place accessories, while the optional equipment tracks can be mounted horizontally. Clinically Flexible There are enough gas (8), data (5), electric (11), plus nurse call options to meet all acuity level needs, including ICU, Med-Surg, PACU, ED/Trauma, Labor and Delivery and patient rooms. Visually Compelling Unit is recessed to take focus off of the headwall and place it where it belongs, on the patient. This also allows for less impact on the space in the room. The Contour® Headwall is available in warm white, with virtually any laminate option to meet your room’s aesthetic needs.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Audible hiss at an outlet, low pipeline pressure alarm on connected equipment, gas smell (O2/N2O).
Root cause: Worn primary/secondary check valve or O-ring in the gas outlet, or a loose pipe union behind the faceplate.
Symptoms: Dead outlet, intermittent power to bedside equipment, tripped breaker.
Root cause: Loosened back-wired terminal, worn receptacle contacts, or overloaded circuit.
Symptoms: Nurse-call station does not register, no network link at the bedside data port.
Root cause: Broken jack, unseated punch-down, or a damaged cable in the raceway.
Symptoms: Rail-mounted devices sag or will not lock, visible track deflection.
Root cause: Mounting fasteners backing out of the wall structure, or track rail worn at the clamp points.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Gas-specific; keyed indexing must match the outlet.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Gas outlet check-valve / O-ring kit (O2, air, vacuum, N2O) Gas-specific; keyed indexing must match the outlet. | Major service part | $$ | Moderate |
| Hospital-grade electrical receptacle | Major service part | $ | Fairly easy |
| Nurse-call / data jack insert | Major service part | $ | Fairly easy |
| Equipment-track rail and clamp hardware | Major service part | $$ | Moderate |
| Faceplate / trim panel | Major service part | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.