Streamlining Workflow. IntelliDrive® XL powered transport helps enable safe and efficient transport of patients throughout the hospital. EXCLUSIVE Powered width expansion reduces time and effort involved in adjusting the bed to accommodate patient size and transport needs. FlexAfoot™ length adjustment enables quick and easy adjustment of bed to fit patient, minimizing migration and need for repositioning. Real-time connectivity, NaviCare® Nurse Call and status boards provide real-time visibility of staff activity, bed status and patient protocols Delivering Safe Patient Care. Programmable Continuous Lateral Rotation Therapy (CLRT) supports protocols for preventing ventilator associated complications. Head-of-bed angle indicator and alert provide responsive monitoring and protocol compliance. 3-mode bed exit supports fall prevention protocols, while the silence feature keeps alarm engaged and ensures alert is set without relying on constant caregiver intervention. Choice of surface options designed to help prevent and treat skin breakdown Enhancing Patient Dignity. Enhanced design, color and styling are consistent with other Hill-Rom standard patient beds. Sturdy siderails and easy access to patient controls and trapeze allow patients to confidently reposition themselves for greater independence and self-sufficiency.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Backrest, knee, height, Trendelenburg or the boost / turn assist will not move from the siderail or caregiver controls.
Root cause: Failed linear actuator, a control-board fault, a control-panel fault, or a pinched / damaged control cable.
Symptoms: Bed will not brake or release, drifts, central brake pedal ineffective.
Root cause: Worn central-brake linkage or castor.
Symptoms: Weight reads implausibly or will not zero, bed-exit alarm false-triggers or fails.
Root cause: Load-cell drift or damage, or a scale-control fault.
Symptoms: Functions and alarms do not work during a power interruption, battery-fault indicator.
Root cause: Aged back-up battery or a charger fault.
Symptoms: Siderail will not latch, rattles, or the CPR release is slow / stiff.
Root cause: Worn siderail latch or the CPR mechanism.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Recalibrate the scale after replacement.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Linear actuator (backrest / knee / height) | Major service part | $$$ | Difficult |
| Central-brake / castor assembly | Major service part | $$ | Difficult |
| Load cell / scale sensor set Recalibrate the scale after replacement. | Major service part | $$$ | Difficult |
| Back-up battery | Periodic replacement | $$ | Fairly easy |
| Siderail control panel / control board | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio