Hillrom · Model Affinity 4
Stow and Go foot section - With this revolutionary maternity bed feature, you are no longer required to lift and remove the foot section. It easily slides into a stored position during delivery and glides back into place afterwards. Battery backup - With this exclusive feature, bed controls continue to function even when the birthing bed is unplugged, facilitating positioning and control during emergency transport. EasyGlide calf supports - Permanently attached to the Affinity 4 Birthing Bed to eliminate misplacing. New single lever lock allows the calf supports to quickly, easily and securely latch into place. OneStep siderails - One handed operation to raise and lower the maternity bed’s siderails. The siderails automatically tuck under the birthing bed when lowered. OneStep foot support - Quick and intuitive to deploy with infinite positioning. Comfortline mattress - New extra thick, gentle air mattress designed to address patient comfort and satisfaction. Tempur-Pedic mattress - The material used in Tempur-Pedic medical mattresses is clinically proven to reduce the incidence of stage II or higher ulcers. Surveys show patients are more comfortable and feel the hospital really cares about patient satisfaction.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Backrest, knee, height, Trendelenburg or the boost / turn assist will not move from the siderail or caregiver controls.
Root cause: Failed linear actuator, a control-board fault, a control-panel fault, or a pinched / damaged control cable.
Symptoms: Bed will not brake or release, drifts, central brake pedal ineffective.
Root cause: Worn central-brake linkage or castor.
Symptoms: Weight reads implausibly or will not zero, bed-exit alarm false-triggers or fails.
Root cause: Load-cell drift or damage, or a scale-control fault.
Symptoms: Functions and alarms do not work during a power interruption, battery-fault indicator.
Root cause: Aged back-up battery or a charger fault.
Symptoms: Siderail will not latch, rattles, or the CPR release is slow / stiff.
Root cause: Worn siderail latch or the CPR mechanism.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Recalibrate the scale after replacement.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Linear actuator (backrest / knee / height) | Major service part | $$$ | Difficult |
| Central-brake / castor assembly | Major service part | $$ | Difficult |
| Load cell / scale sensor set Recalibrate the scale after replacement. | Major service part | $$$ | Difficult |
| Back-up battery | Periodic replacement | $$ | Fairly easy |
| Siderail control panel / control board | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio