Haemonetics · Model Cell Saver 5+
Advanced Cell Saver technology. Built-in SmartSuction technology. Sophisticated data management technologies. User-friendly touch screen.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Bowl vibration or noise, speed error, wash aborts.
Root cause: Bowl not seated in the chuck, worn centrifuge chuck/latch, or an out-of-balance bowl set.
Symptoms: Bowl over- or under-fills, wash starts early or late, poor product hematocrit.
Root cause: Dirty or failed optical sensor at the bowl, or air in the line.
Symptoms: "State fault", pump speed/direction error, self-loading pump will not seat tubing.
Root cause: Peristaltic pump rollers/door, valve actuator, or a CPU/safety-system disagreement.
Symptoms: Unit dies when unplugged, low-battery early.
Root cause: Aged battery pack or charger fault.
Symptoms: Fluid trapped in or leaking from the bowl core during processing.
Root cause: Defective 70/125/225 ml bowl set - subject of Haemonetics recalls.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Single-procedure consumable; check the recall status of the lot.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Disposable bowl / reservoir set Single-procedure consumable; check the recall status of the lot. | Consumable | $ | Easy |
| Optical line sensor / bowl optics | Major service part | $$ | Moderate |
| Peristaltic pump roller assembly | Periodic replacement | $$ | Moderate |
| Centrifuge chuck / latch | Major service part | $$$ | Difficult |
| Battery pack | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Potential to develop leaks at the inner core of Cell Saver 5/5+ and Cell Saver Elite/Elite +125ml and 225ml bowl sets. Which could result in fluid being trapped inside the bowl core, causing a long empty device error message with the possibility of incomplete washing.
Haemonetics has received isolated reports that in some cases the 70ml centrifuge bowl may leak during cell salvage processing.
Device emits radiofrequency which exceed the IEC 60601-1-2 International Standard
Disposable Set for Autologous Blood recovery may leak saline at the Y-connector