Haemonetics · Model cardioPAT
Separation technology. Dynamic Disk® consistently delivers 70–80% hematocrit. Reservoir. QuickConnect reservoir supports “collect first” approach. Partial collection and return. Processes up to 2 liters per hour and as little as. 5 milliliters of RBCs. Safety alerts LCD display provides operational and safety alerts. Suction Onboard suction with SmartSuction Harmony System. RBC bags with integrated microaggregate filter. Rechargeable battery facilitates blood collection and data storage during patient transport.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Bowl vibration or noise, speed error, wash aborts.
Root cause: Bowl not seated in the chuck, worn centrifuge chuck, or an out-of-balance bowl set.
Symptoms: Bowl over- or under-fills, wash starts early or late, poor product hematocrit.
Root cause: Dirty or failed optical sensor at the bowl, or air in the line.
Symptoms: "State fault", pump speed/direction error, self-loading pump will not seat tubing.
Root cause: Worn pump rollers/door, a valve actuator, or a CPU/safety-system disagreement.
Symptoms: Unit dies when unplugged, low-battery early.
Root cause: Aged battery pack or charger fault.
Symptoms: Fluid trapped in or leaking from the bowl core during processing.
Root cause: Defective bowl set - subject of Haemonetics Cell Saver / Elite recalls; check the lot.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Single-procedure; check recall status of the lot.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Disposable bowl / reservoir set Single-procedure; check recall status of the lot. | Consumable | $ | Easy |
| Optical line sensor / bowl optics | Major service part | $$ | Moderate |
| Peristaltic pump roller assembly | Periodic replacement | $$ | Moderate |
| Centrifuge chuck / latch | Major service part | $$$ | Difficult |
| Battery pack | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Synced live from the FDA openFDA API.
Updated Operations Manual for the deployment of the spill collection system and cleaning
Update Instructions for Use for the CardioPAT® Manual for Leak Detection. User may mis-interpret the air-leak rate displayed by the cardioPAT device when making a determination about chest-drain removal, which could result in the patient experiencing a pneumothorax.
Labeling (additional) use instructions: to clarify proper position of cardioPAT device and post-op line to prevent fluid build up from patient's chest
Blood clotting of patient post operative while on the cardioPAT Cardiovascular Perioperative Autotransfusion System
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.