Getinge Disinfection AB · Model 46-5
WASHER DISINFECTOR 46-5 S+W D, S+W D indicating single door machine with dryer. Cat no was 46-5-402. From March 2019: 46-5-502.
Intended use: A mains electricity (AC-powered) unit designed to wash (remove dirt/stains), decontaminate (remove organic debris), and disinfect (destroy microorganisms) reusable surgical instruments, bowls, anaesthesia equipment components, and some surgical footwear. Most instruments, especially those used invasively or in a sterile field will require sterilization after this process. It is typically located in rooms and central supply departments designed to handle soiled instruments/utensils. It typically includes internal baskets/racks in which the utensils are placed to ensure circulation of the heated water/detergents to all surfaces.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Pressure loss, long cycle, steam leak.
Root cause: Aged rubber gasket.
Symptoms: No steam, low temp, cycle abort.
Root cause: Failed heating element or scale buildup.
Symptoms: Poor air removal, non-condensable gas fault.
Root cause: Worn vanes/seals.
Symptoms: Scaling, conductivity alarm.
Root cause: RO/DI system failure.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable.
Most-replaced wear item.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Vacuum pump & seals | Periodic replacement | $$$ | Difficult |
| Chamber temperature sensor | Periodic replacement | $$ | Moderate |
| RO/DI feed cartridge Consumable. | Frequent wear item | $ | Easy |
| Door gasket / seal Most-replaced wear item. | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.
There is potential for a cart with disinfected goods to unload without the trolley being docked. Disinfected goods could end up falling to the floor causing risk to nearby users and damage to goods and equipment.
Getinge Disinfection AB has initiated a Field Correction due to the potential for unauthenticated remote access to the product over the network.
Getinge Disinfection AB received complaints regarding sediment residuals in the manifold of the washer and disinfector of Getinge 46-Series (CM302) with PACS 300/350 control. After performing detailed root cause analysis, the firm has identified the problem as insufficient water flow throughout the manifold due to formation of air pockets where sedimentation may occur.