GE HealthCare · Model Universal PAC
Carrying case.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Agent analyser shows the output does not match the dial beyond the vaporiser's stated tolerance.
Root cause: The vaporiser is out of calibration - recalibration is done only at an authorised service centre.
Symptoms: Machine leak test fails with this vaporiser fitted, agent smell.
Root cause: Worn back-bar O-rings / mounting seal, or the vaporiser not fully locked / Selectatec latched onto the manifold.
Symptoms: Will not fill (keyed or funnel filler), spills, or overfills; sight glass hard to read.
Root cause: Worn filler seal / valve, or the wrong agent bottle adapter.
Symptoms: Dial stiff, will not turn past the 0 / T interlock, or does not stay at a setting; Selectatec interlock not releasing.
Root cause: Worn dial detent / interlock or an internal control-cone fault.
Symptoms: Grossly high output for a period after the vaporiser was tipped or moved.
Root cause: Agent carried into the bypass channel after tipping - purge at high flow before clinical use.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Vaporizers are calibrated only at an authorised service centre.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Back-bar / mounting O-ring & seal kit | Periodic replacement | $ | Fairly easy |
| Filler seal / valve | Periodic replacement | $ | Fairly easy |
| Agent-specific fill adapter | Major service part | $ | Easy |
| Vaporiser recalibration (authorised service centre) Vaporizers are calibrated only at an authorised service centre. | Frequent wear item | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
Under certain workflows, patient information shown in the viewer may not match the images displayed on Centricity Universal Viewer Zero Footprint Client (ZFP) with versions ZFP v6.0: SP11.6, SP11.6.0.1 and SP11.6.1 on systems with Patient Timeline installed,
There is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be identified which could allow a malicious actor with these credentials to access the system and potentially manipulate patient data.
GE HealthCare has become aware of an issue in Centricity Universal Viewer Zero Footprint Client (ZFP) where there is a potential security vulnerability which could allow a malicious actor to access the system and potentially manipulate patient data.
GE HealthCare has become aware of an issue in Centricity Universal Viewer Zero Footprint Client (ZFP) version v6.0 SP11.x where the latest report is not shown by default to the user. The issue may occur when a study has multiple DICOM structured reports (i.e. preliminary, final, and/or addendum) stored in different series.
Footprint Client (ZFP) versions v6.0 SP9.x and SP10.x where the latest addended report is not shown by default to the user. The issue occurs when the addendum is created on the same day as the original report but at a different time, and only when launched on the following browsers. " Internet Explorer 10 (IE10) " Internet Explorer 11 (IE11) " Firefox.
When using Universal Viewer or Centricity Universal Viewer with Workflow Manager (WFM), if a third-party reporting application is used to launch exams, it could display information on a different patient than WFM and the Viewer.
GE HealthCare has become aware of an issue where information from two different patients can be mismatched when correcting patient or study information.
Inaccurate distance measurements with magnified projection X-ray images.
Images from the Centricity PACS-IW with Universal Viewer and Centricity Universal Viewer may be missing when a system parameter MapRoute is set to a value greater than 1.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio