GE HealthCare · Model Signa HDxt 1.5T
Built on the High Definition platform users know and trust, Signa HDxt delivers more definitive diagnosis in the most challenging exams. Four new clinical applications offer unrivaled performance enhanced with anatomy-specific components that take accuracy and certainty to a new level. Truly gapless data in 30% less time with Cube 3D Isotropic Imaging. Get a full diagnostic story - any way users slice it with a single 3D volume scan. Outstanding fat suppression, every time with IDEAL. A revolutionary technique for uniform fat suppression, especially in challenging areas. Just one scan delivers four images: fat only, water only, in-phase, and out-of-phase. Uncompromised Image Quality in a Single Breathold High-resolution, in-phase and out-of-phase imaging - in just one breath-hold using 3D Dual Echo for Signa HDxt. Investment protection with the Signa Continuum. In 23 years, no GE customer has been left behind in MR technology. Signa HDxt leverages GE Magnet Technology that is built to last (1.5T and 3.0T), applications and hardware with our 25 year Continuum so they're never out of date, and a variety of newly compatible tables only from GE.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Low signal-to-noise, shading, or a dead region in images; coil not detected or fails the coil check.
Root cause: Damaged RF coil element, a coil-cable / connector fault, or a preamp fault.
Symptoms: Geometric distortion, gradient-amplifier faults, acoustic change, over-temperature trips.
Root cause: Gradient-amplifier fault, gradient-coil cooling problem, or a gradient-coil fault.
Symptoms: Rising helium boil-off, cold-head noise or stall, He level dropping, compressor alarms.
Root cause: Cold-head at end of life, a helium-compressor fault, or a chilled-water problem.
Symptoms: System-cooling alarms, gradient or RF over-temperature, scans aborting.
Root cause: Facility chilled-water supply, the heat exchanger, or internal pumps.
Symptoms: Table will not dock, drive or index; landmarking errors.
Root cause: Table drive, docking sensor, or encoder fault.
Symptoms: Recon errors, host hangs, images will not reach PACS.
Root cause: Recon or host computer fault, or the network / storage.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Planned replacement; He boil-off rises as it wears.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| RF receive coil (exchange) | Major service part | $$$$ | Moderate |
| Cold head Planned replacement; He boil-off rises as it wears. | Periodic replacement | $$$$ | Difficult |
| Helium compressor / adsorber | Major service part | $$$$ | Difficult |
| Gradient amplifier module | Major service part | $$$$ | Expert-level |
| Chilled-water pump / heat exchanger | Major service part | $$$ | Difficult |
| Patient table drive / encoder | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.