GE HealthCare · Model Ohmeda 4400
• Microprocessor Control • 20 Segment Output Control • Digital Apgar Timer • Powered Elevating Base • Pre-Heat Mode • Dovetail Rail Mounting System • Large bed accommodates larger infants easily • Bed height adjusts to comfortable working level • Advanced microprocessor system continuously monitor patient and machine status • Easy patient access from any location around unit • Heater Power: 550 Watts ±5% at maximum % power setting • Examination light Luminance: 100 (nominal) foot candles at center of mattress
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: No heat, overheat alarm.
Root cause: Element or thermal fuse failure.
Symptoms: Setpoint error, alarm.
Root cause: Sensor aging.
Symptoms: Uneven warming, hotspot.
Root cause: Blower wear.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable per patient.
Common R&R.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Blower fan | Periodic replacement | $$ | Fairly easy |
| Temperature probe Consumable per patient. | Periodic replacement | $ | Easy |
| Heating element / thermal fuse Common R&R. | Periodic replacement | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →