GE HealthCare · Model MAC 800
Compact and easy to carry, the MAC 800 weighs only 3kg with the battery and is designed with an integrated carrying handle. Lithium ion battery provides two hours of continuous operation or 250 standard patient reports. Fast battery recharge (about four hours) minimizes system downtime. Easy-to-use keypad with T9 SMS mobile phone keypad design and quickaccess function keys. Hospital-grade keypad allows for easy cleaning. 7-inch, high-resolution (800 x 480) color TFT screen. Digital onscreen preview lets you view your ECG prior to printing, saving paper and avoiding costly repeat exams. Tested replaceable accessories with multi-link lead wires, fixed lead wires, disposable electrodes and reusable electrodes. Marquette 12SL analysis program. Hookup Advisor™ informs clinicians if there is poor waveform quality during ECG recording. Acute Cardiac Ischemia Time-Intensive Predictive Instrument (ACI-TIPI) included in the Marquette 12SL predicta the probability of cardiac ischemia, as well as factoring in gender-specific criteria to help physicians more readily diagnose heart abnormalities in women. RR analysis* also is available to detect hidden patterns of underlying heart rate variability (HRV) and to measure cardiac RR intervals.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Baseline wander, muscle or AC noise, "lead fail", poor 12-lead quality.
Root cause: Poor skin prep, dried or lifting electrodes, or a worn patient cable / lead set.
Symptoms: Acquisition hangs, no interpretation, or a self-test failure.
Root cause: Acquisition-module fault or a software issue.
Symptoms: No or faint report, paper jam, misregistered grid.
Root cause: Thermal print head, paper handling, or (older units) the stylus.
Symptoms: Short battery run time, will not hold charge, will not power on.
Root cause: Aged battery pack.
Symptoms: ECGs will not transmit to the ECG-management system / EMR.
Root cause: Network / Wi-Fi / modem configuration or the communication module.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient cable / 10-lead set | Frequent wear item | $$ | Easy |
| Acquisition module | Major service part | $$$ | Moderate |
| Thermal print head / paper | Periodic replacement | $ | Fairly easy |
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Communication / modem module | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.