Fujifilm · Model FDR Go
Compact chassis size and low tube lockdown allow easy driving visibility and maneuverability in even the tightest spaces. Large touchscreen display with Fujifilm’s latest Dynamic Visualization™ image processing, which automatically optimizes images for greater diagnostic visibility. Compatible with all FDR D-EVO wireless detectors available in GOS and CsI, in standard 14x17” or full field-of-view 17x17” sizes, and the unique wireless 24x30cm all featuring Fujifilm’s patented ISS technology for exceptional image quality at lower dose. Dual, independent rear-wheel drive provides ultra-sensitive steering for smooth, quiet traveling and easy maneuvering in tight spaces. “Inching” controls at the tube head slowly move the system forward or backward for precise positioning at the bedside without needing to return to the drive handle. Long- and tall-reach telescopic arm features full angulations for easy and precise tube positioning. System battery power lasts 4-5 hours and features emergency reserve mode for additional exposures and travel after low battery warning. System can be operated wireless or with wired Ethernet connection utilizing standard DICOM connectivity and its ruggedized hard drive holds up to ~4,000 images. RIS shortcut button lets you close studies remotely from the mobile console.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Exposure inhibited, kV / mAs error, images under- or over-exposed.
Root cause: HV generator / capacitor fault, exposure-switch fault, tube fault, or a thermal interlock.
Symptoms: Unit will not drive, drives weakly, or a drive fault code.
Root cause: Drive battery, drive motor / controller, or the throttle handle.
Symptoms: Few exposures per charge, will not reach full charge, unit dies mid-round.
Root cause: Aged exposure / drive battery bank or a charger fault.
Symptoms: Tube arm drifts, will not extend / rotate, or a brake will not hold.
Root cause: Worn arm bearings, counterbalance, or brakes.
Symptoms: Panel not detected, images not received, console errors.
Root cause: Wireless detector, its battery, or the acquisition PC.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Battery bank (exposure / drive) | Periodic replacement | $$$ | Moderate |
| HV generator / capacitor module | Major service part | $$$$ | Expert-level |
| Drive motor / controller | Major service part | $$$ | Difficult |
| Column arm brake / bearing kit | Major service part | $$ | Moderate |
| Wireless detector / battery (DR) | Periodic replacement | $$$$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.