Fresenius · Model 2008T
Uses conductivity sensors to measure dialysate sodium concentration pre- and post-dialyzer to determine effective dialyzer urea clearance Measures and predicts adequacy goal attainment Alerts clinician if target Kt/V will not be met Facilitates early intervention during the treatment to achieve adequate therapy On-screen, step-by-step instruction for monthly access flow measurement when used in conjunction with the Twister® reverse-flow bloodline Trending of access flow which permits intervention before access failure Auto Flow feature allows the user to set the dialysate flow to either 1.5 times or 2 times the blood flow in steps of 100 mL/min.3 Durinq idle periods, the 2008T machine will automatically run dialysate flow at 300 mL/min to conserve dialysate or it may be manually set to as low as 0 mL/min. Ensures the most economical use of dialysate while maintaining an optimal dialysis dose - potential savings of up to 30 percent with no impact on Kt/V attainment.4 Four preset profiles as well as four facility-customizable profiles are available. An automated method of changing the concentration of dialysate sodium in accordance with the physician's prescription.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Conductivity or temperature out of range, dialysate-preparation alarm, treatment will not start.
Root cause: Concentrate connection / supply, a heater fault, a conductivity-cell fault, or a proportioning-pump / valve fault.
Symptoms: Pump will not run at set speed, occlusion or flow alarms, tubing walking in the raceway.
Root cause: Worn pump rotor / rollers or a motor / control fault.
Symptoms: Venous / arterial / TMP pressure implausible, air-detector false alarms or fails to detect.
Root cause: Transducer-protector wetting, a pressure-sensor fault, or an air-detector fault.
Symptoms: Fluid-balance / UF alarms, weight loss inaccurate, balancing-chamber leak.
Root cause: Balancing-chamber diaphragm / valve wear or a flow-sensor fault.
Symptoms: Heat or chemical disinfection will not complete, residual disinfectant test fails.
Root cause: Heater, a dosing / rinse valve, or a conductivity-verification fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Blood / dialysate pump rotor & rollers | Periodic replacement | $$ | Moderate |
| Conductivity cell / temperature sensor | Periodic replacement | $$ | Moderate |
| Pressure sensors / transducer protectors | Periodic replacement | $$ | Fairly easy |
| Balancing-chamber diaphragm / valve kit | Periodic replacement | $$$ | Difficult |
| Proportioning pump / valves | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.