B. Braun · Model Dialog Plus
B. Braun Dialog® PLUS Hemodialysis System B. Braun's new Dialog® Plus Hemodialysis System incorporates sophisticated medical technology while offering convenient handling in an appealing and ergonomic design. This state-of-the-art hemodialysis system achieves an ideal balance for ease of operation with its flat screen technology and innovative use of icons on a color touch screen, a flat monitor display to facilitate fast and easy handling, on-screen HELP functions that can be accessed instantly and a wide selection of pre-programmed and user-definable treatment profiles. Dialog Plus Hemodialysis System's unique On-Line Technical Data and Advanced Technical Service Mode provide the technician with a complete on-screen technical overview available while in the dialysis mode and during maintenance, thus reducing troubleshooting and maintenance downtime. Graphic display capabilities providing an overview of the subsystem components, and configuration screens which allow the technician to set default treatment values and modes, simplify servicing problems and can help to shorten technician training time. 270 degree rotatable color screen allows for easy viewing from all angles.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Conductivity or temperature out of range, dialysate-preparation alarm, treatment will not start.
Root cause: Concentrate connection / supply, a heater fault, a conductivity-cell fault, or a proportioning-pump / valve fault.
Symptoms: Pump will not run at set speed, occlusion or flow alarms, tubing walking in the raceway.
Root cause: Worn pump rotor / rollers or a motor / control fault.
Symptoms: Venous / arterial / TMP pressure implausible, air-detector false alarms or fails to detect.
Root cause: Transducer-protector wetting, a pressure-sensor fault, or an air-detector fault.
Symptoms: Fluid-balance / UF alarms, weight loss inaccurate, balancing-chamber leak.
Root cause: Balancing-chamber diaphragm / valve wear or a flow-sensor fault.
Symptoms: Heat or chemical disinfection will not complete, residual disinfectant test fails.
Root cause: Heater, a dosing / rinse valve, or a conductivity-verification fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Blood / dialysate pump rotor & rollers | Periodic replacement | $$ | Moderate |
| Conductivity cell / temperature sensor | Periodic replacement | $$ | Moderate |
| Pressure sensors / transducer protectors | Periodic replacement | $$ | Fairly easy |
| Balancing-chamber diaphragm / valve kit | Periodic replacement | $$$ | Difficult |
| Proportioning pump / valves | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
Extracted directly from this unit's own service manual.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.