Fanem · Model Duetto 2386
Touchscreen monitor Buttons for conversion between incubator and radiant infant warmer mode Pedal for converting from incubator mode to radiant infant warmer X-ray cassette drawer Audible and visual safety alarms External access devices to minimize contact with the patient and the internal environment of the incubator Independent air and oxygen inlets, with filters Information on the monitor can be set for different languages Indication of date, time and trend curves with history Indication of the proportional power level of the heating element Set points and alarm limits saved in the memory Automatic auto-test system In-bed scale Programmable servo control humidity system Programmable O2 servo control system Adjustable height ergonomic support Auxiliary skin temperature sensor – T2 Drawer for storing material Programmable servo control temperature and oxygen concentration system Redundant safety system, with auto shutdown of the heating system in the event of high temperatures Pole for infusion pump and IV pole
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: No heat, overheat alarm.
Root cause: Element or thermal fuse failure.
Symptoms: Setpoint error, alarm.
Root cause: Sensor aging.
Symptoms: Uneven warming, hotspot.
Root cause: Blower wear.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Common R&R.
Consumable per patient.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Heating element / thermal fuse Common R&R. | Periodic replacement | $$$ | Moderate |
| Blower fan | Periodic replacement | $$ | Fairly easy |
| Temperature probe Consumable per patient. | Periodic replacement | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →