Fanem · Model Diapump R2D2
Aspiration with smooth adjustment from 0 to 22 in.-Hg (560 mm-Hg) Dual protection system that cuts off aspiration when the bottle is full Flow of 0 - 24 L/min Highly lit and easy-to-view LED panel Easy-to-clean membrane keypad Full-bottle alarm with automatic shutdown device Operates with or without a pedal, and with an indicator on the panel Silenced alarm indicator.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Gauge low or zero, poor suction at the catheter.
Root cause: Worn pump diaphragm / vanes, a hydrophobic filter not seated or clogged, or a canister seal leak.
Symptoms: Regulator knob has no effect.
Root cause: Worn vacuum regulator.
Symptoms: Nothing happens on power-up, or it stops after a period.
Root cause: Switch / cord fault, or a thermal cut-out on a worn motor.
Symptoms: Fluid or aerosol reaching the pump / exhaust.
Root cause: Overflow float or hydrophobic filter fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Pump diaphragm / vane rebuild kit | Periodic replacement | $$ | Difficult |
| Hydrophobic / bacterial filter | Consumable | $ | Easy |
| Collection canister and lid seal | Frequent wear item | $ | Easy |
| Vacuum regulator | Major service part | $$ | Moderate |
| Vacuum gauge | Major service part | $ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.