Fanem · Model Diapump Colibri DPM 40+
Microprocessed control panel, with LED display; Motor assembly protection box, made of resistant plastic, with IPX1 protection Portable, with handle handles, mounted on metal structure furniture (pedestal) supported on four 3 ”diameter rotating casters, with brake Oil-free, requiring no lubricants Vacuometer for measuring negative pressure calibrated to 30 inHg (760 mmHg) Rotary knob for adjusting the pressure intensity Air filter at the outlet of the aspirated air to the environment Internal battery system (UPS), with autonomy of up to 40 minutes Optional audiovisual alarm for full bottle, with automatic suction shutdown Low battery voltage alarm and battery operation indication Aspiration flow: 30LPM Intelligent level sensor recognition, with auto programming when turning on the equipment Electronic level detector system incorporated in the bottles; Inlet suction: -80 kPa / -600 mmHg (+/- 5%)
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Weak or no vacuum at the patient port, slow to pull down, will not reach rated vacuum.
Root cause: Worn pump diaphragm / vanes, an air leak in the lid / tubing / overflow float, or a blocked filter.
Symptoms: Suction drops suddenly, fluid near the pump, filter wet.
Root cause: Overflow shut-off triggered or a saturated / clogged bacteria filter.
Symptoms: Pump noisy, hot, runs slow, or trips its thermal cut-out.
Root cause: Worn motor bearings or pump head.
Symptoms: Short run time, will not power on off mains.
Root cause: Aged battery.
Symptoms: Gauge inaccurate or vacuum will not regulate.
Root cause: Gauge or regulator wear.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Pump diaphragm / vane kit | Periodic replacement | $$ | Moderate |
| Hydrophobic / bacteria filter | Consumable | $ | Easy |
| Collection-jar lid seals & float | Frequent wear item | $ | Easy |
| Motor / pump head | Major service part | $$$ | Difficult |
| Battery (portable) | Periodic replacement | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio