EPPENDORF · Model Centrifuge 5804 / 5804 R
• Interchangeable swingbucket and high-speed fixedangle rotors • Spin up to 14,000 rpm, or 20,800 x g • Low-profile 11.2 in/28 cm access height for easy sample loading • Automatic rotor recognition senses rotor type to set maximum allowable speed - ensures safety • Fast, simple programming of time, temperature (refrigerated models), speed/ g-force and radius value • 34 user-defined programs • Adjustable-speed Short Spin button • Ten acceleration and braking rates • Speed, g-force and radius correction values can be entered • Values can be changed during centrifugation • Temperature range from –9 °C to 40°C, with Fast Temp function for rapid cooling of the centrifuge chamber • Standby cooling maintains temperature when not in use • Speed can be set from 200 rpm to maximum speed in increments of 10 rpm • Short Spin button • Selectable program memory for up to 34 individual user programs • Ten acceleration and deceleration rates for sensitive sample material
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Aspiration error, short sample, carryover.
Root cause: Clot/crystal obstruction.
Symptoms: Vibration, loud noise, stop error.
Root cause: Worn bearing or unbalanced rotor.
Symptoms: Clogs, high blank, QC fail.
Root cause: Pump tube wear, air in line.
Symptoms: Incubator out of range.
Root cause: Failed heater/fan or sensor.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable, replace proactively.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Peristaltic pump tubing Consumable, replace proactively. | Frequent wear item | $ | Easy |
| Xenon / LED light source | Periodic replacement | $$$ | Moderate |
| Centrifuge rotor & bearings | Major service part | $$$ | Difficult |
| Incubator heater / fan | Periodic replacement | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.