EnRaf Nonius · Model Radarmed 650
Continuous or pulsed output Effortless transition between continuous and pulsed therapy The parameters are conveniently arranged on the illuminated display. The parameter settings remain visible on the display during treatment. Free memory for the storage of treatment protocols (only on Radarmed 950+)
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Output power display stays at "000" with power commanded, no therapeutic heating at the emitter.
Root cause: Magnetron at end of life, a high-voltage supply fault, or the emitter-detection circuit not seeing the emitter.
Symptoms: Fan does not start on power-up, rattles, or airflow blocked; magnetron overheats and output cuts.
Root cause: Failed or obstructed cooling fan, or a blocked air intake around the magnetron.
Symptoms: H.F. output does not switch off at the end of the set time, no end-of-treatment signal.
Root cause: Faulty mechanical treatment timer or its switch.
Symptoms: Output present at the unit but weak or no field at the emitter, intermittent with cable movement.
Root cause: Damaged emitter or coax cable / connector.
Symptoms: Three-segment LED display dark or wrong reading.
Root cause: Display or control fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
High-voltage service; discharge and follow the manual.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Magnetron High-voltage service; discharge and follow the manual. | Periodic replacement | $$$ | Difficult |
| Cooling fan | Periodic replacement | $ | Fairly easy |
| Mechanical treatment timer | Major service part | $$ | Moderate |
| Microwave emitter / coax cable | Major service part | $$$ | Fairly easy |
| High-voltage supply components (capacitor/diode) | Major service part | $$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.