Elmed · Model ESU 100
Three RF-solid state circuits. One for unipolar cutting and coagulation in the MICRO or MACRO modes, one for bipolar coagulation, and one for BI-POL CUTTING ESU 100 E M/M-B only). Accepts all laparoscopic or pelviscopic electrosurgical instruments. Single pedal footswitch operated, with low voltage switching circuit. Patient safety circuit. Automatic, low voltage sentry, alarms and deactivates the unipolar output in case of a defect in the connections for the patient plate circuit. Optionally, our Return Electrode Monitor System is available, which also will alarm and deactivate the unipolar output in case of poor patient plate contact. Change-over switch for unipolar or bipolar operation, defeats the patient plate alarm in the bipolar mode. Visual and audible RF-indicators. L.E.D. Bar-graph meter for unipolar wattage. Low voltage bipolar generator provides reliable coagulation with all available bipolar forceps for laparoscopy, pelviscopy and microsurgery. Musical coagulation monitor for the bipolar mode provides both audible and visual indication of the degree of tissue coagulation, particularly for bipolar tubal sterilization (Not available on the 100 L). SMART-O-MATIC, automatic compensation circuit provides the desired output for all electrodes or tissue conditions. The automatic range control, which is divided into ten steps, requires no readjustment within the set range. The SMART-O-MATIC operates in both the unipolar and bipolar modes. Output intensity may be precisely regulated within the full range of the dial setting. However, investigations have demonstrated 20 W at 100 Ohms achieves optimal results, therefore, we have incorporated a preset selection and 20 W range preselect switch (Not available on 100 L). All of these ELMED 100 series Electrosurgical Units are designed as table top units. A specially designed cart (List No. 52-5067) is available for these described units and other endoscopic equipment.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: No cut / coag effect, "high impedance" or output-error message, output shuts down.
Root cause: Poor electrode contact or a dry field, a damaged handpiece / cable, a return-electrode (REM/NESSY) fault, or an output-stage fault.
Symptoms: REM / NESSY / plate alarm, monopolar output inhibited.
Root cause: Poorly applied or dried split dispersive pad, a pad-cable fault, or the contact-quality-monitor circuit.
Symptoms: Bipolar or sealing output weak, seal cycle will not complete, instrument not recognised.
Root cause: Bipolar cable / instrument fault or the sealing output stage.
Symptoms: Output power or the REM circuit outside tolerance on the analyser.
Root cause: Output board or REM circuit needs recalibration - anti-static handling required.
Symptoms: Output will not activate from the pedal or the handpiece.
Root cause: Footswitch or handswitch / cable fault.
Symptoms: Unresponsive touchscreen, will not power on or boot, fan / thermal alarm.
Root cause: Display, power supply, control board, or a cooling-fan fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Recalibration required after replacement; anti-static handling.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Reusable instrument / handpiece cables | Periodic replacement | $$ | Easy |
| Split dispersive (return) pads | Consumable | $ | Easy |
| Footswitch | Periodic replacement | $$ | Fairly easy |
| Bipolar / sealing cable | Periodic replacement | $$ | Easy |
| RF output board Recalibration required after replacement; anti-static handling. | Major service part | $$$$ | Difficult |
| Cooling fan | Major service part | $ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.