Edwards Lifesciences · Model EV1000
Choice - provides the choice of the parameters you want to view and how you want to view them. Clarity - color-based indicators communicate patient status at a glance, and visual clinical support screens allow for immediate recognition and increased understanding of rapidly changing clinical situations for improved decision making. Clinical Support - presenting the physiologic status of the patient in an entirely new and meaningful way enables you to focus on your patient.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: CO / CI blank or dashed, poor arterial waveform quality flagged.
Root cause: FloTrac / VolumeView sensor not zeroed or connected, a damaged pressure cable, or a poor arterial line.
Symptoms: Pressure trace noisy, absent, or will not zero.
Root cause: Worn or damaged reusable pressure cable / connector.
Symptoms: Venous oximetry will not calibrate, values drift.
Root cause: In-vitro / in-vivo calibration not done, or an optical-module / cable fault.
Symptoms: Unresponsive touchscreen, will not run on battery, shuts down.
Root cause: Display fault or an aged internal battery.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Single-patient disposable.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Reusable pressure cable | Periodic replacement | $$ | Easy |
| FloTrac / VolumeView sensor Single-patient disposable. | Consumable | $ | Easy |
| Oximetry cable / optical module | Periodic replacement | $$$ | Fairly easy |
| Internal battery | Periodic replacement | $$ | Fairly easy |
| Touchscreen assembly | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio