DXP · Model Axyia 6
Autoclave Axyia 6, a class S and N autoclave, was designed specifically to meet the needs of the professional beauty sector. Characterized by an elegant and compact design, it weighs less than 20 kg and boast a chamber with a 6 litre capacity. Just like practices, beauty centres, barber shops and hairdressers also need to manage the high contamination risk caused by the presence of germs, bacteria and organic residue on their equipment day after day. European regulations have established clear standards for a safe and traceable sterilization. Axyia 6 makes it simple to effectively comply with this obligation thanks to a reliable technology built using the best materials and characterized by an excellent price/quality ratio. The functioning of the autoclave is curated in every detail to guarantee high-performance and reliability: air is expelled through the thermodynamic system, creating a vacuum inside the chamber in a quick and silent manner while the advanced microprocessor cross-checks pressure and temperature guaranteeing high-quality saturated vapour. Axyia 6 is based on two fundamental values – ease of use and sturdiness. 4 programmes and automatic functions such as troubleshooting, self-diagnosis, night-cycle, cycle counter and remote control make it extremely easy to use. The autoclave fully complies with standard EN13060 and is made entirely of stainless steel. The chamber is made of seamless stainless steel and each pressure vessel is a single piece.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Cycle aborts, temperature or pressure not reached / held, failed BI or chemical indicator.
Root cause: Heater / element fault, a door-gasket leak, a steam-generation or water-fill fault, a blocked chamber drain / strainer, or a temperature / pressure-sensor fault.
Symptoms: Steam leak at the door, door will not seal, lock or unlock, pressure will not build.
Root cause: Worn door gasket or a door-lock / interlock fault.
Symptoms: Instruments / packs come out wet.
Root cause: Drying-phase vacuum / heater fault, an overloaded chamber, or a clogged drain.
Symptoms: Low-water alarm, mineral scale, cloudy chamber, poor steam quality.
Root cause: Empty or scaled reservoir, a fill-valve / pump fault, or use of non-distilled water.
Symptoms: Failed Bowie-Dick / air-removal test, slow come-up, air pockets.
Root cause: Blocked steam trap, a vacuum-pump / ejector fault, or a check-valve fault.
Symptoms: No or faint cycle printout.
Root cause: Thermal print head or paper.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Door gasket | Periodic replacement | $$ | Fairly easy |
| Heating element / steam-generator element | Major service part | $$$ | Difficult |
| Chamber / feedwater strainer & fill valve | Periodic replacement | $$ | Moderate |
| Temperature / pressure sensor | Major service part | $$ | Moderate |
| Steam trap / solenoid & check valves | Periodic replacement | $$ | Moderate |
| Thermal print head / paper | Periodic replacement | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio