Dräger · Model Infinity
Touchscreen, plus keyboard and mouse navigation. Electronic “white board” for entering notes using standard text entry. Clinical screens to support advanced analysis including ST, VentCentral® ventilation management, trends and calipers. 2-hour Full and Event Disclosure storage standard. Up to 72-hours of Full Disclosure waveforms including EEG. Up to 16 waveforms and 1,000 arrhythmia or alarm events stored per patient. HL7 interface for patient admission and laboratory values. Alarm adjustment at patient monitor or central station. Sophisticated server includes quiet fan operation, integrated and external speakers, and RAID option. Ability to view up to 600 patients across the Infinity Network. Intranet access to view information conveniently where you need it. Ability to export ECG data to Holter systems and Infinity MegaCare® for long-term storage.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: A bed shows no ECG or dropped beats at the central station, noisy trace, frequent leads-off.
Root cause: Worn transmitter lead set, a weak transmitter battery, or antenna / receiver coverage.
Symptoms: Dropout in specific areas of the unit, multiple beds affected in one zone.
Root cause: Antenna cabling, a splitter / amplifier, or a receiver-module fault.
Symptoms: Central station hangs, will not admit, alarms not annunciating, display frozen.
Root cause: Central-station PC / server, the application, or the network link.
Symptoms: No strip on alarm or on demand.
Root cause: Central recorder / network printer fault.
Symptoms: Timestamps wrong or drifting, full-disclosure gaps.
Root cause: Network time source or connectivity.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Telemetry transmitter lead set | Frequent wear item | $$ | Easy |
| Transmitter battery (per pack) | Consumable | $ | Easy |
| Receiver module / antenna components | Major service part | $$$ | Difficult |
| Central-station PC / server | Major service part | $$$$ | Difficult |
| Network printer / recorder | Major service part | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.