Datrend · Model Infutest 2000 Series E
Unmatched Accuracy: Better than 1% for flow, pressure, and volume measurements - ideal for critical-care pump models Multi-Pump Testing: Evaluate up to 4 IV channels simultaneously for high-efficiency biomedical workflows Versatile Compatibility: Supports single-rate, dual-rate, and PCA (Patient-Controlled Analgesia) infusion devices Intelligent Test Automation: Use manual or autosequence testing via intuitive front-panel keypad and graphical interface Advanced Features: SynchroStart™: Automatically captures startup flow dynamics for pulsatile pumps End-of-Infusion Analysis (EIA): Accurately calculates VTBI when infusion ends Interoperable: Interfaces seamlessly with ES601Plus Safety Analyzer, medTester 5000C, and other automated test platforms (Generated by AI)
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Readings out of tolerance on verification check.
Root cause: Reference component aging.
Symptoms: Won't power on, unexpected shutdown mid-test.
Root cause: Aged internal battery or PSU.
Symptoms: Intermittent or no reading through a specific lead.
Root cause: Repeated connect/disconnect cycles.
Symptoms: Blank, frozen, or garbled display.
Root cause: Display module or driver board aging.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Send out for annual re-certification.
High-turnover; keep spares.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Display module | Major service part | $$$ | Difficult |
| Calibration reference module Send out for annual re-certification. | Major service part | $$$$ | Difficult |
| Rechargeable battery pack | Periodic replacement | $$ | Fairly easy |
| Test lead set High-turnover; keep spares. | Periodic replacement | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio →