Cynosure · Model Icon
• Skintel Melanin Reader: A revolutionary device that determines the average melanin density of skin prior to treatment • Advanced Contact CoolingTM: Cooling maintains temperature at 5° Celsius during treatments for enhanced comfort and skin protection. • Various Microlenses for customized treatments: -XFTM Microlens: High-speed, non-ablative fractional resurfacing with increased surface coverage typically allows for a full face, one-pass treatment in less than 15 minutes. -XDTM Microlens: Non-ablative fractional resurfacing that uses point compression and a unique lens design to allow for a greater depth of penetration. • AccuSpectrumTM: Dual filters minimize unnecessary epidermal exposure, enhancing client comfort. • Photon RecyclingTM: Maximizes treatment results by reducing energy loss. • Smooth PulseTM: Proprietary light delivery system that allows for the usage of greater amounts of energy. • Power: Delivering a maximum effective energy of over 350 joules, Icon is the most powerful laser and optimized light aesthetic platform.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Reduced output power, inconsistent treatment fluence.
Root cause: Normal diode/gain-medium wear over lamp/diode hours.
Symptoms: Overheat shutdown mid-treatment, error code on startup.
Root cause: Coolant pump, TEC element, or filter degradation.
Symptoms: Uneven beam delivery, connector arcing, or no output at the handpiece.
Root cause: Fiber or articulated-arm optics degraded by flex/heat cycling.
Symptoms: Won't fire, safety interlock fault on the display.
Root cause: Door/footswitch interlock or shutter mechanism failure.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Core consumable; major cost driver.
High-turnover; keep spares.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Footswitch / interlock assembly | Major service part | $$ | Fairly easy |
| Laser diode / flashlamp module Core consumable; major cost driver. | Periodic replacement | $$$$ | Difficult |
| Cooling system (pump / TEC / filter) | Periodic replacement | $$$ | Moderate |
| Handpiece / delivery fiber assembly High-turnover; keep spares. | Periodic replacement | $$$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.