Capintec · Model CRC-25R Dose Calibrator
USB and RS232 port SD flash card software upgrade USB printer capability Chamber plug-and-play capability Interactive intelligent remote display Both remote and chamber can be placed 100 feet from the readout unit Selection of Nuclide and Daily Test can be done with the remote On screen display of Nuclide Name, Number, Activity, Unit of Measure and Calibration Number Large character, high visibility display with automatic backlighting Over 80 Nuclides with half-lives in memory Automatic zero and background subtraction Built-in dose calibration, quality control and self diagnostics Includes a pre-set key for F-18 measurements Compatible with Nuclear Medicine Management Systems via USB Optional printer for full size NRC records and patient labels for syringes and vials. Inventory control Dose table calculations Enhanced software includes linearity and geometry protocol Radiochromatorgraphy QC package Built in geometry testing Standard 2 year warranty.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Daily constancy or quarterly accuracy/linearity checks trend out, activity readings off vs the reference source.
Root cause: Ionization-chamber gas or pressure change, electrometer drift, or a contaminated well/liner.
Symptoms: Background will not zero, noisy readings at low activity.
Root cause: Electrometer leakage, chamber contamination, or nearby radioactive sources.
Symptoms: Readout unit does not see the remote or the chamber, no activity value.
Root cause: The long interconnect cable (up to ~30 m) or a connector fault.
Symptoms: No NRC-record printout or syringe/vial labels.
Root cause: Printer cable, paper, or the optional printer.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Capintec service; recalibration required.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Ionization chamber Capintec service; recalibration required. | Major service part | $$$$ | Difficult |
| Electrometer board | Major service part | $$$ | Difficult |
| Chamber / remote interconnect cable | Major service part | $$ | Fairly easy |
| Well liner | Periodic replacement | $ | Easy |
| Optional printer / paper | Periodic replacement | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
Software error in Rev. 2.01 and Rev 2.02 for Capintec CRC 25PET
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio