Canon · Model Vantage Fortian 1.5T
Remote monitoring and seamless patient handling with mobile Tablet UX Assist patient positioning and coil set-up with Ceiling Camera and Intelligent Monitor Efficiently plan with ForeSee View and automate routine sequences with Auto Scan Assist Easy to clean gloss coating aids disinfection requirements Advanced intelligent Clear-IQ Engine (AiCE) helps you to see through the noise to deliver clear, sharp and distinct images Compressed SPEEDER for 2D and 3D, FAST 3D and parallel imaging scan technologies to reduce scan time Robust imaging with motion correction and reduced distortion technologies mART, Iterative Motion Correction (IMC), RDC-DWI Advanced diagnosis with Olea/Vitrea Small footprint and low energy consumption to help you minimize your operational costs MR Theater helps relax patients with a virtual immersive experience Pianissimo Σ and Pianissimo Zen technology delivers whisper quiet scanning Short magnet and 71cm wide bore offers an open patient friendly environment.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Low signal-to-noise, shading, or a dead region in images; coil not detected or fails the coil check.
Root cause: Damaged RF coil element, a coil-cable / connector fault, or a preamp fault.
Symptoms: Geometric distortion, gradient-amplifier faults, acoustic change, over-temperature trips.
Root cause: Gradient-amplifier fault, gradient-coil cooling problem, or a gradient-coil fault.
Symptoms: Rising helium boil-off, cold-head noise or stall, He level dropping, compressor alarms.
Root cause: Cold-head at end of life, a helium-compressor fault, or a chilled-water problem.
Symptoms: System-cooling alarms, gradient or RF over-temperature, scans aborting.
Root cause: Facility chilled-water supply, the heat exchanger, or internal pumps.
Symptoms: Table will not dock, drive or index; landmarking errors.
Root cause: Table drive, docking sensor, or encoder fault.
Symptoms: Recon errors, host hangs, images will not reach PACS.
Root cause: Recon or host computer fault, or the network / storage.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Planned replacement; He boil-off rises as it wears.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| RF receive coil (exchange) | Major service part | $$$$ | Moderate |
| Cold head Planned replacement; He boil-off rises as it wears. | Periodic replacement | $$$$ | Difficult |
| Helium compressor / adsorber | Major service part | $$$$ | Difficult |
| Gradient amplifier module | Major service part | $$$$ | Expert-level |
| Chilled-water pump / heat exchanger | Major service part | $$$ | Difficult |
| Patient table drive / encoder | Major service part | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.