Canon · Model RadPRO Mobile 40kW
Canon Flat Panel Detectors The RadPRO Mobile 40kW Digital X-ray System is compatible with the Canon portable and wireless digital radiography detectors. Depending on your organization's needs, each flat panel detector series has either a Gadolinium Oxysulfide or Cesium Iodide scintillator, helping to produce high resolution images that may be previewed within 3-5 seconds. Single or Dual Panel Solution Single detector and dual panel detector configurations are available. Mobile Radiography Room Rather than bring the patient to the radiography room, you can bring the radiography room to the patient with this mobile solution. Patients with limited mobility, need assistance at bedside, or in wheelchairs can easily be imaged. For added technologist's convenience, 17" inch touch-screen with integrated generator controls allows the technologist to simply select the study being performed and automatically program the generator to the correct technique. With storage areas for all that a technologist needs during a mobile examination, the RadPRO Mobile 40kW Digital X-ray System is an invaluable resource for the technologist. Maneuvering and Positioning With a mere press of the handle, large 17 inch wheels that are powered by 2 independent drive motors easily tackles typical facility obstacles. With power-assisted pressure sensitive steering, the RadPRO Mobile 40kW Digital X-ray System is simple to control and drive while minimizing the amount of force required by the technologist. The telescoping tube arm provides 6 additional inches in length to provide flexible positioning management at a patient's bedside. Equipped with precision positioning buttons, a technologist can easily adjust the mobile unit from the handles of the tube arm if needed. Generator/Power The high frequency generator provides up to 150kVp, 500mA, and 500mAs which, when paired with the standard 300KHU x-ray tube, is powerful enough to shoot through most patients, including bariatric. With 2 battery systems, the RadPRO Mobile 40kW Digital X-ray System provides independent power for exposure control and the drive system reducing power drain and allowing for increased productivity.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Exposure inhibited, kV / mAs error, images under- or over-exposed.
Root cause: HV generator / capacitor fault, exposure-switch fault, tube fault, or a thermal interlock.
Symptoms: Unit will not drive, drives weakly, or a drive fault code.
Root cause: Drive battery, drive motor / controller, or the throttle handle.
Symptoms: Few exposures per charge, will not reach full charge, unit dies mid-round.
Root cause: Aged exposure / drive battery bank or a charger fault.
Symptoms: Tube arm drifts, will not extend / rotate, or a brake will not hold.
Root cause: Worn arm bearings, counterbalance, or brakes.
Symptoms: Panel not detected, images not received, console errors.
Root cause: Wireless detector, its battery, or the acquisition PC.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Battery bank (exposure / drive) | Periodic replacement | $$$ | Moderate |
| HV generator / capacitor module | Major service part | $$$$ | Expert-level |
| Drive motor / controller | Major service part | $$$ | Difficult |
| Column arm brake / bearing kit | Major service part | $$ | Moderate |
| Wireless detector / battery (DR) | Periodic replacement | $$$$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio