Burdick · Model Vision Premier Holter
Advanced capabilities, including automatic categorization of heartbeats, tachogram review, and full-disclosure viewing. Automatic categorization of Holter information into unique morphology forms, including normal, ventricular, paced, SVE, unknown, and artifact. System algorithms that automatically identify cardiac abnormalities and compile them into easy-to-read reports. 7-second diagnostic strip labeling for each cardiac abnormality. Includes classification, time of occurrence, heart rate, and heartbeat annotations. Additional ECG-strip (EKG-strip) information can be inserted at any point. On-screen displays for global or individual episode review or edit. Automatic ST changes analyzes in three channels. Adjusts the analysis for different heart rates. Software with a category-specific histogram to review the frequency and severity of an arrhythmia. Tachogram review feature to identify, review, and note atrial fibrillation/flutter. On-screen calipers for easy and precise measurements. Can superimpose single- or multi-beat modes complemented with a 30-second full-disclosure review and RR interval plot. Multiple ECG scanning modes (paging, superimposition, or retrospective review) for reviewing data, capturing arrhythmias, and compiling a final report. Collect and analyze up to 48 hours of 3-channel ECG recordings. Optional remote mode to download and review heart rate and ECG data anytime, anywhere.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Baseline wander, muscle or AC noise, "lead fail", poor 12-lead quality.
Root cause: Poor skin prep, dried or lifting electrodes, or a worn patient cable / lead set.
Symptoms: Acquisition hangs, no interpretation, or a self-test failure.
Root cause: Acquisition-module fault or a software issue.
Symptoms: No or faint report, paper jam, misregistered grid.
Root cause: Thermal print head, paper handling, or (older units) the stylus.
Symptoms: Short battery run time, will not hold charge, will not power on.
Root cause: Aged battery pack.
Symptoms: ECGs will not transmit to the ECG-management system / EMR.
Root cause: Network / Wi-Fi / modem configuration or the communication module.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Patient cable / 10-lead set | Frequent wear item | $$ | Easy |
| Acquisition module | Major service part | $$$ | Moderate |
| Thermal print head / paper | Periodic replacement | $ | Fairly easy |
| Battery pack | Periodic replacement | $$ | Fairly easy |
| Communication / modem module | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio