BPL Medical Technologies · Model E-CUBE 15 Platinum
21.5″ Wide LED Monitor with Super Responsive Touch Screen 10.4" touch screen with user friendly UI and customizable presets Greater sensitivity, Wider bandwidth, Higher resolution Motorized-height adjustment of control panel Enhanced visualization of structures, clear depth perception, and color sensitivity Comprehensive cardiac measurement and analysis package Unparalleled image quality in 2D mode, Color Flow, and Doppler modes Crystal Signature™ transducers to help increase penetration and deliver a high definition resolution Volume Master™ - ALPINION’s intuitive 3D/4D software package in convex and endocavity transducers offers fast rendering speed and outstanding rendering quality for diagnosis Live HQ™ - The combination of qualified 3D/4D transducers with strong performance reliability and geometrical accuracy enables ALPINION’s Live HQ™ software to render a realistic fetal shape Cube Strain™ & Stress Echo-advanced cardiac software package for accurate diagnosis Needle Vision ™ & Needle Vision Plus: to enhance the visibility of needle tip and shaft, making the procedure easy Cube View ™ - World’s first Real time image sharing software to improve your patient throughput and give you a diagnostic edge Advanced 3D/4D rendering engine (Volume Master™ and LIVE HQ™) offering excellent details of fetus High performance, lightweight Crystal Signature™ volume convex transducer for volume acquisitions Wide linear transducer to evaluate breast abnormalities User-oriented system and transducer design fulfills the ultimate comfort for the users as well as patients.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Noisy, low-contrast or non-uniform image, missing sector, drop-out that follows the system not the probe.
Root cause: Front-end / beamformer board fault, a probe-port (connector) board fault, or a system-board fault.
Symptoms: A probe works on one port but not another, or is not detected at all.
Root cause: Worn probe-port connector or a scan-connector board fault.
Symptoms: Dead system, hangs at boot, or powers off under load.
Root cause: Power supply / DC converter, the system battery, or a system-board fault.
Symptoms: Unresponsive TGC sliders or keys, distorted or blank display.
Root cause: Control-panel board, touchscreen, or the monitor.
Symptoms: Thermal-shutdown warning, fan roar, hot exhaust.
Root cause: Failed or dust-clogged cooling fan or blocked air path.
Symptoms: Studies will not send to PACS, images will not archive, worklist empty.
Root cause: Network configuration or a storage-drive fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Front-end / beamformer board | Major service part | $$$$ | Difficult |
| Probe-port / scan-connector board | Major service part | $$$ | Difficult |
| Power supply / DC converter | Major service part | $$$ | Difficult |
| Control-panel / touchscreen assembly | Major service part | $$$ | Moderate |
| Cooling fan | Major service part | $ | Moderate |
| System hard drive | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.