Bovie Bipolar (BBp) - a pulsing coagulation for quick and powerful coagulation of tissue Auto Bipolar Intuitive Technology - immediate and simplified tissue coagulation with minimal tissue damage Surgical Memory Technology - recalls the last used power and mode Adjustable Power During Activation - during all activations, the power can be adjusted up or down in variable steps, giving surgeons more control at the surgical site BovieNEM™ – This return electrode contact quality monitoring system is designed to minimize patient burns at the return electrode site. Bovie FDFS™ – The tissue sensing technology measure tissue 5 million times a second and adjusts to varying impedances 5000 times a second for accurate energy delivery.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Poor cut/coag, sparking, low power.
Root cause: Burned/eroded blade or pencil.
Symptoms: Fractured tip, loose component, debris.
Root cause: Metal fatigue, improper reprocessing, or misuse.
Symptoms: Residue, bioburden, discoloration.
Root cause: Inadequate cleaning cycle or damaged tray.
Symptoms: Slow or drifting positioning.
Root cause: Cylinder seal or actuator wear.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Single-use; expiry critical.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Return electrode (patient plate) Single-use; expiry critical. | Consumable | $ | Easy |
| Foot switch assembly | Periodic replacement | $$ | Fairly easy |
| OR light LED module | Major service part | $$$ | Difficult |
| Table hydraulic actuator | Periodic replacement | $$$ | Difficult |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.