Biocare · Model iE 300
High resolution 5 inch color LCD screen. Simultaneous 12 leads acquisition . Digital ECG technology with 24 bit resolution 8000 samples / sec. Frequency response 0.05 Hz ~ 250 Hz. Automatic ECG diagnosis for adults, children and newborns. Simultaneous recording Functions: AUTO, Manual and Rhythm. Thermal printer with 80 mm paper. Ability to print in the following formats: 1 X 12, 1 X 12 + 1R, 3 X 4, 3 X 4 + 1R, 3/2. Printing speeds (5, 6.25,10,12.5,25,50) mm / s. Visual and audible alarms for: low battery charge, blank paper tray or disconnected cables. Built-in lithium battery. Battery charge level indicator. Easy operation with one-touch shortcut keys Up to 300 seconds of parsing - facilitates arrhythmia detection. Pest repellent filters: EMG, Baseline wander and AC. Automatic and manual sensitivity adjustment in 7 levels (1.25, 2.5, 5, 10, 20, 10/5, 20/10) mm / mV & Auto Gain. Implantable pacemaker detection. Freeze screen. Internal digital data storage memory with a capacity of 3000 files. Two USB ports for expanding the internal memory and connecting the barcode scanner. Network port (LAN) for sending data to a central station. Weight of device with battery = 1.3 kg. CE mark & ISO.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Flat or noisy ECG, frequent lead-off alarms.
Root cause: Worn electrodes, loose leads, or dry gel.
Symptoms: No pleth, SpO2 drops to --- or false low.
Root cause: LED/photodiode aging or cable damage.
Symptoms: Cuff fails to inflate, erroneous BP, long cycle.
Root cause: Pump/valve wear or cuff leak.
Symptoms: Short backup time, battery alarm on transport.
Root cause: Aged backup battery.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Consumable per patient.
High-turnover; keep inventory.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| NIBP cuff & hose Consumable per patient. | Periodic replacement | $ | Easy |
| Rechargeable backup battery | Periodic replacement | $$ | Fairly easy |
| Display / touch assembly | Periodic replacement | $$$ | Difficult |
| ECG patient cable & lead set High-turnover; keep inventory. | Frequent wear item | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.