Bio-Rad Laboratories, Inc. · Model CFX96 Deep Well
Deep wells for large volumes Detection of up to 5 targets per well, plus a channel dedicated to singleplex FRET Protocol autowriter that generates an optimal protocol for your reaction components Thermal gradient that identifies optimal annealing temperature in a single run.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Wells at the edge vs centre give different Ct, failed thermal verification, ramp errors.
Root cause: Peltier block ageing / a failed Peltier zone, a heated-lid fault, or a fan / heatsink fault.
Symptoms: High background, low signal, dye / ROI or pure-dye calibration fails, poor amplification curves.
Root cause: Excitation LED / lamp ageing, optics contamination, or an out-of-date calibration.
Symptoms: Evaporation, condensation on the seal, inconsistent replicates, lid-pressure error.
Root cause: Heated-lid heater or the lid-pressure mechanism / seal.
Symptoms: Instrument not detected, runs will not start or export, firmware mismatch.
Root cause: USB / network link, driver, or the control PC.
Symptoms: Slow ramp-down, over-temperature warning, loud fan.
Root cause: Failed or clogged cooling fan.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Peltier sample block | Major service part | $$$$ | Difficult |
| Excitation LED / lamp module | Periodic replacement | $$$ | Moderate |
| Heated-lid heater / seal | Major service part | $$ | Moderate |
| Cooling fan | Periodic replacement | $ | Fairly easy |
| Dye calibration kit | Frequent wear item | $$ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
The connector of the power supply was capable of fitting other components provided with the system. However, the voltage output of the reader power supply is incompatible with the other component and, if incorrectly connected, could result in irreparable damage to the system.
The value for the ANA Cutoff and ANA Positive Controls will decrease overtime resulting in an unusually elevated positivity rate for patient samples tested.
From BioLab, for work done with this kind of equipment.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.