Beckman Coulter · Model Iris iChemVELOCITY
Analyzes up to 210 samples per hour 300 strip load capacity Up to 10,000 patient record storage capacity Prevents use of expired components Ready-to-use consumables provides fewer errors when reconstituting Load samples, press START and walk away for effortless operation Continuous strip loading capability while system is running Ascorbic acid detection informs of potential interference with key chemistry assays to deliver more clinical information Easily recall QC or patient results with user-friendly software Single action strip loading process On-board QC traceability Clear visibility into strip chamber for easy monitoring of strip supply.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Strip results flagged or biased, QC out of range, colour blocks misread.
Root cause: Expired or mishandled strips, a dirty / aged optics head, or a calibration issue.
Symptoms: Strip not transported, jams, or read at the wrong position.
Root cause: Worn strip-transport belt / tray or a position-sensor fault.
Symptoms: Repeated optics errors, non-uniform reads across the pads.
Root cause: Contaminated or aged reflectance LEDs / optics.
Symptoms: Poor particle images, flow-cell or camera errors.
Root cause: Flow-cell contamination, a pump fault, or the camera / focus.
Symptoms: No printout, results not sent to the LIS.
Root cause: Printer or the interface configuration.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Strip transport belt / tray | Periodic replacement | $$ | Moderate |
| Optics / reflectance head | Major service part | $$ | Moderate |
| Calibration strips / QC | Frequent wear item | $ | Easy |
| Sediment flow cell / pump tubing (if fitted) | Frequent wear item | $$ | Moderate |
| Printer / paper | Periodic replacement | $ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
A defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computers metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the devices CD-ROM.
The Leukocyte and Compensation pads were switched on some vials of lot number 7212269M of the iChem VELOCITY Strips.
Customers have reported incidents where two sets of results report the same Specimen Identifier (Specimen ID) with different results, different Medical Record Number (MRN) and different patient identification (demographics).
The probe misalignment or bending has the potential to lead to delay in reporting results or incorrect patient results.