Symptoms: Lift / lower or traverse does not respond to the hand control.
Root cause: Discharged or mis-inserted battery, an active emergency stop, an obstruction or overload, or a faulty hand control / cable.
Symptoms: Will not raise or lower, runs one direction only, noisy or slow, strap frays.
Root cause: Worn lift actuator / motor, or a frayed / twisted lifting strap or chain.
Symptoms: Short run time, frequent low-battery, unit dies mid-transfer.
Root cause: Aged battery pack or a charger / charging-contact fault.
Symptoms: Carriage will not move along the rail, or binds.
Root cause: Worn carriage wheels, a rail obstruction, or the traverse motor.
Symptoms: Lift stops with an overload indication, slack-strap cut-off active, or emergency lowering does not work.
Root cause: Load over the rated limit, the slack-strap sensor, or the emergency-lowering mechanism.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Inspect before every use; retire on any fraying.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Rechargeable battery pack | Periodic replacement | $$ | Easy |
| Hand control | Periodic replacement | $$ | Easy |
| Lifting strap / chain Inspect before every use; retire on any fraying. | Periodic replacement | $$ | Fairly easy |
| Lift actuator / motor assembly | Major service part | $$$ | Difficult |
| Rail-carriage wheel set (ceiling) | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Synced live from the FDA openFDA API.
The upper portion of the hooks on the spreader bar are too flexible for holding the loops of the sling. The loop of the sling could potentially come out of the hooks if positioned in the upper portion of the hooks.
The ceiling lift may not stop lateral movement after releasing the left (or right) action button on a ceiling lift equipped with an infrared hand control.
The Kwiktrak Gate stopper may fail and allow the ceiling lift to pass through to the end of the track and fall to the floor.