American Medical Systems · Model Aura XP
• Precise Cutting and Coagulation - StarPulse™ Technology precisely limits the depth of penetration for increased control, while providing excellent hemostasis, vaporization, and cutting power. • Unique StarPulse Technology - Allows you to deliver a pulse width as short as 1 millisecond. • 532 nm Pulsed Wavelength - KTP energy is highly absorbed by oxyhemoglobin, delivering precise coagulation while leaving the surrounding tissue unaffected. • Compact Design and Quiet Operation - Perfect for in-office and small procedure rooms and can be placed on a countertop or cart. Its small, compact design is perfect for the hospital, office or ambulatory surgery center. • Easy and Convenient Plug and Play Operation - Standard Electrical 120vac (20amp) plug and play operation enables easy movement from one room to another and expedites procedure time. • Solid-state Reliability - Reliable, solid-state construction sets a new standard for high performance with easy maintenance.
Failure patterns typical of this equipment class - general BMET reference, not verified against this exact unit's own service history unless flagged. Severity is a category-level estimate.
Symptoms: Reduced output power, inconsistent treatment fluence.
Root cause: Normal diode/gain-medium wear over lamp/diode hours.
Symptoms: Overheat shutdown mid-treatment, error code on startup.
Root cause: Coolant pump, TEC element, or filter degradation.
Symptoms: Uneven beam delivery, connector arcing, or no output at the handpiece.
Root cause: Fiber or articulated-arm optics degraded by flex/heat cycling.
Symptoms: Won't fire, safety interlock fault on the display.
Root cause: Door/footswitch interlock or shutter mechanism failure.
The parts a BMET typically services on this class of equipment. Service class and relative cost are general engineering estimates for the category, not figures from this unit's documentation. Part numbers appear only when pulled from this unit's own service manual. Always confirm intervals and pricing against the manufacturer.
Core consumable; major cost driver.
High-turnover; keep spares.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Laser diode / flashlamp module Core consumable; major cost driver. | Periodic replacement | $$$$ | Difficult |
| Cooling system (pump / TEC / filter) | Periodic replacement | $$$ | Moderate |
| Handpiece / delivery fiber assembly High-turnover; keep spares. | Periodic replacement | $$$ | Fairly easy |
| Footswitch / interlock assembly | Major service part | $$ | Fairly easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A structured starting point derived from the failure modes above. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Synced live from the FDA openFDA API.
The product size shown on the carton labeling does not match the size on the product inside.
Mis-labeling: The Dura-II 2 cm Universal Tips and the Dura-II 4 cm Distal Tips were incorrectly labeled with the opposite labels.
Overheating of the Cryoablation Therapy Disposable Probe (CU-1), could cause a burn injury to a patient or user in certain circumstances.