Error Codes
Extracted directly from this unit's own service manual.
DIAGNOSTIC 0When diagnostic 0 is selected, the Capacitor is charged up to its charge level, and the capacitor charge as read through the A/D circuitry is displayed. The second line of display shows the signal noise (should be less than 1.5 Volts)
DIAGNOSTIC 1Displays the Energy Monitor voltage. The second line of the display shows the signal noise
DIAGNOSTIC 2Displays the voltage from the Temperature Sensor, ≈2.9 VDC @ room temperature. The second line of the display shows the signal noise
DIAGNOSTIC 3Displays the voltage from the HED which monitors the shutter position. This voltage is ≈7.5 VDC with the shutter open, ≈3.5 VDC with the shutter closed. The second line of the display shows the signal noise. Press the TEST FIRE button to see the voltage level which the computer recognizes as “closed” boundary. Press the AIMING button to see the voltage level that the computer recognizes as the bou
DIAGNOSTIC 4not used
DIAGNOSTIC 5/6/7Displays the %VDC (Diag 5), +15VDC (Diag 6), and -15VDC (Diag 7) voltages as measured by the A/D circuit
DIAGNOSTIC 8Used to select the language for the display. Use the TEST FIRE and AIMING buttons to scroll back and forth through the language selections
DIAGNOSTIC 9/10/11Display Q Fire Volts, Cap volts, and Double Fire volts. In Service Diagnostic Mode, AIMING/TEST FIRE increase/decrease the stored value. Normally these values are determined and entered during Automatic Laser Calibration
DIAGNOSTIC 12Displays the Double Fire count. TEST FIRE/AIMING decrease/increase the stored value in Service Diagnostic mode. In Service Diagnostic mode the snubber can be turned off/on with the Service Switch (leave snubber on)
DIAGNOSTIC 13/14Diagnostic 13 is used to display one of the 16 stored energy monitor calibration values - move from one value to another with the TEST FIRE and AIMING buttons. Values 1 through 6 are average voltage out of the energy monitor at each of the six wheel settings. Values 9 through 14 are average mj at each of the six wheel settings. Use Service Diagnostic 14 to run the Energy Monitor Calibration
DIAGNOSTIC 15/16Diagnostic 15 displays the watchdog minimum time. Diagnostic 16 displays the watchdog maximum time. Press TEST FIRE to see the spec value. Press TEST FIRE and AIMING at the same time to see the difference between spec and actual. For Rev J software - min 3.5 msec (±1 msec) max. 8.5 msec (±2 msec)
DIAGNOSTIC 17Displays the “Total Hours.” This value should not be reset
DIAGNOSTIC 18Displays “Total Shots.” Reset when the YAG head is replaced. Reset by pressing the Service and TEST FIRE buttons at the same time
DIAGNOSTIC 19Use to initialize the EAROM when new software EAROM is installed. Push the Service and RESET buttons at the same tome to initialize the EAROM
1AttenSwitch Both poles open/closed Bad attenuator switch, broken wire in head harness, or U9 on the Digital PCB. Bad ribbon cable System to Digital PCB's
2ZeroCross No Zero Crossing Trigger On System PCB; U2, During self test only. U3, or Q2. Bad ribbon cable to CPU
3OpenConnect Connector unplugged Digital PCB U9. EC2, P20, P21, P16, P7 Ribbon cable
41 ms. Timer Wrong amt. time in test System PCB;U8, U5, U6, EC2 on backwards
5A/D Convert Convcmpt hasn’t been seen Analog PCB; U14, U16, U17, or CPU PCB
6ShutterFail Shutter position incorrect Shutter stuck open. Enter shutter voltage in SERVICE mode, diagnostic #3, broken wires in Head Harness
7FootSwitch Footswitch is shorted U9 Digital PCB, EC2
83.5 ms 1 Shot Outside ±1 ms Digital PCB, watch dog, U10
98.5 ms 1 Shot Outside ± 2 ms Digital PCB, watch dog, U10
10Check Rom 0 CPU; bad prom #1,U21
11Check Rom 1 CPU; bad prom #2, U20
12Check Rom 2 CPU; bad prom #3,U19
13Check Rom 3 CPU; bad prom #4,U18
14RAM Check RAM bad at initialization CPU; bad ram, U13
15EAROM Resp. EAROM not responding CPU; bad EErom, (EAROM not initialized) U12 or initialize/re-initialize ROM
16EAROM Time EAROM stays busy CPU; bad EAROM, U12, re-initialize, cycle on/off
17+5 Volts A/D reading chg >220mv +5 volts changed or ±15 VDC changed ±15VDC Supply bad
18Laser Volts Not chgd. to val. in 1.5 sec YPSESAFE on System PCB; Q6, Q8, or U8; High Voltage PS; Q7
19Watch Dog W.dog T.O. or miss task CPU bad; or Digital PCB U1, U2, U7 orU4 Turbo Q3
20Power Fail Missed Zero xng Twice Zero Crossing circuit on System PCB; U2, U3, or Q2 (Fault 22 occurs in pre REV F software only. Later revisions display ??? after two misfires.)
21Laser Fire Two misfires, no signal System PCB; U4, or head not firing,CPU, or EEROM U12
22Laser Det. Two missed detects, System PCB; U1, signal read Q1 or U7
23Double Fire Double fire detected Real double fire, or noise on System PCB, U7, pin 11, check diag 12 - recal E mon
24IR Filter High signal, no detect Not enough IR filters in CAL in Energy Monitor housing. Check TP11 on System PCB
25Prog. Lost It got to a place with no code CPU bad, Xfmr fuse (.5 A)
26EARom CkSum EAROM Check Sum Error CPU; EAROM, U12 bad.
27LasSupHi Over Voltage for 1.5 seconds HV Power Supply; cap over-charged, bad Q6 and CR1, U1, Q4 on Turbo P/S
282 Zero Cross No Zero Crossing Trigger On System PCB; U5 or U3 During self test only on Digital PCB
29OpenConnect Connector unplugged U8 Digitial, P7 from Digital, EC2, P16, P20, P21
315 A/D Convert Conversion Complete not seen Analog PCB; U14, U16, U17, or CPU PCB
326 ShutterFail Shutter position not correct Shutter stuck open. Enter shutter voltage in SERVICE mode, diagnostic #3, possible broken wires
359 8.5 ms 1 Shot Outside ± 2 ms Digital PCB, watch dog U10
4014 RAM Check RAM bad at initialization CPU; bad RAM U6 re-initialize
4115 EAROM Resp. EAROM not responding CPU; bad EEROM, U5 cycle on/off
4216 EAROM Time EAROM stays busy CPU; bad EEROM, U5
4418 Laser Volts Not chgd to value in 1.5 sec. System PCB TP3
4620 HVPS Latch HVPS latch did not clear System PCB;U10
4721 Power Fail Missed Zero Crossing Twice Zero Crossing circuit on System PCB; U5 or U3
5025 IR Filter High signal, no detect in CAL Not enough IR filters in Energy Monitor housing. Check TP8 on System PCB, U2, U14
5126 Prog. Lost It got to a place with no code CPU bad, Xfmr fuse (.5 A)
5227 EAROM CkSum EAROM Check Sum Error CPU; EAROM bad, U5
5328 LasSupHi Over Voltage for 1.5 seconds HV Power Supply; cap over-charged, bad SCR1, U2
5429 LasNoSnub Not Snubbing Not snubbing cap. HVPS; or no Snubber signal from System PCB, TP3 and TP2, U2
5531 NUMBER 31 NMI not disabled System PCB; Rev F retrofit performed incorrectly (Q4 not removed)