American Diagnostic Corporation · Model Pro's Combo II SR
Combines ADC's ProsphygTM 768 with an AdscopeTM Sprague 641 and deluxe carry case Contemporary black enamel 300mmHg manometer Adcuff™ calibrated nylon cuff with proprietary Size Guide™ marking system prevents mis-cuffing Chrome-plated deflation valve Complete aneroid, matching case, and color-matched Sprague stethoscope Extra large, color-matched, dual pocket carry case, measures 9” x 5” x 2” Not made with natural rubber latex Assembled, inspected, and packaged in the U.S.A. from foreign components Meets or exceeds ANSI/AAMI/ISO 81060-1.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Reads high or low vs a reference, needle off zero at rest.
Root cause: Wear or a shock event to the movement.
Symptoms: Pressure will not hold, slow steady drop on the gauge.
Root cause: Perforated bladder, cracked tubing, or a leaking connector.
Symptoms: Bulb will not build pressure, deflates too fast or sticks.
Root cause: Worn air-release valve or bulb one-way valve.
Symptoms: Cracked inflation bulb or a leaking connector.
Root cause: Age and handling.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Aneroid gauge / movement | Major service part | $$ | Fairly easy |
| Cuff and bladder | Periodic replacement | $ | Easy |
| Air-release valve | Periodic replacement | $ | Easy |
| Inflation bulb | Periodic replacement | $ | Easy |
| Connecting tubing / connector | Periodic replacement | $ | Easy |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio