Aloka · Model SSD-4000
The SSD-4000 PureHD system uses key features of the Aloka ProSound Series Quint Frequency Imaging allows up to five different frequency settings for a single transducer. Harmonic Echo is a technology that utilizes second harmonics, which include much less side lobe artifact and noise. This also helps the imaging of difficult to exam patients, such as the aged or the obese. An integrated DICOM-compatible image management provided by Aloka's Data Management Subsystem (iDMS) that includes DICOM work-list capability. Dual Dynamic Display (DDD) simultaneously displays two B-mode images (with and without color flow) to easily show the relationship between the morphological information and the blood flow of the region of interest. Multi Beam Processing offers exceptionally high frame rates for optimal performance in studies requiring dynamic imaging of patient morphology. The SSD-4000 can provide a frame rate as high as 500 frames/second or more (transducer dependent). Even with the general abdominal transducer UST-9123, a frame rate up to 286 frames/second is possible. Real-time Free Angular M-Mode technology provides superior imaging regardless of positioning. It makes the exam fast and accurate, while providing exceptional image quality.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Noisy, low-contrast or non-uniform image, missing sector, drop-out that follows the system not the probe.
Root cause: Front-end / beamformer board fault, a probe-port (connector) board fault, or a system-board fault.
Symptoms: A probe works on one port but not another, or is not detected at all.
Root cause: Worn probe-port connector or a scan-connector board fault.
Symptoms: Dead system, hangs at boot, or powers off under load.
Root cause: Power supply / DC converter, the system battery, or a system-board fault.
Symptoms: Unresponsive TGC sliders or keys, distorted or blank display.
Root cause: Control-panel board, touchscreen, or the monitor.
Symptoms: Thermal-shutdown warning, fan roar, hot exhaust.
Root cause: Failed or dust-clogged cooling fan or blocked air path.
Symptoms: Studies will not send to PACS, images will not archive, worklist empty.
Root cause: Network configuration or a storage-drive fault.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Front-end / beamformer board | Major service part | $$$$ | Difficult |
| Probe-port / scan-connector board | Major service part | $$$ | Difficult |
| Power supply / DC converter | Major service part | $$$ | Difficult |
| Control-panel / touchscreen assembly | Major service part | $$$ | Moderate |
| Cooling fan | Major service part | $ | Moderate |
| System hard drive | Major service part | $$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.