Alco · Model Electric Bariatric Plus
Quiet, smooth operation/Easy-to-use Hand Pendant electrically controls Head/Foot/Contour positioning and overall height of the bed frame for maximum comfort and versatility for the patient and caregiver. Clip-on pendant mounts on bed rail, within easy reach to operate all adjustments. Variable Comfort Chair and Cardiac Chair (with Trendelenburg/Reverse Trendelenburg staff control option) adjustability helps provide added comfort and relaxation positioning for the resident and facilitates caregiver accessibility. Cardiac Chair positioning is beneficial for residents with heart and respiratory problems that find the semi-upright position brings comfort and relief by elevating the torso from the waist up, which provides relief to the lungs, increases circulation. Trendelenburg/Reverse Trendelenburg capable positioning with Staff Control Option is ideal for rehab and sub acute care, provides therapeutic benefits that can aid in circulation and ease discomfort. Edema foot ratchet elevates the foot end of the bed. 4 heavy duty Euro style roll at any height steer & locking casters provide easy maneuverability and secure lock positioning. Adjustable wall saver minimizes distance between wall to bed frame to prevent damage with/without optional Trapeze. Adjustable mattress retainer securely supports mattress position for 80'', 84", 88" length. Built in push button width expansion from 39" to 54" width. All Steel frame and no sag slat deck construction provides superior strength, durability, mattress support and is easy to clean.
Compiled for this model from its own documentation and established engineering knowledge for this equipment class - not from a verified service record. Confirm against the manufacturer's service manual.
Symptoms: Backrest, knee, height, Trendelenburg or the boost / turn assist will not move from the siderail or caregiver controls.
Root cause: Failed linear actuator, a control-board fault, a control-panel fault, or a pinched / damaged control cable.
Symptoms: Bed will not brake or release, drifts, central brake pedal ineffective.
Root cause: Worn central-brake linkage or castor.
Symptoms: Weight reads implausibly or will not zero, bed-exit alarm false-triggers or fails.
Root cause: Load-cell drift or damage, or a scale-control fault.
Symptoms: Functions and alarms do not work during a power interruption, battery-fault indicator.
Root cause: Aged back-up battery or a charger fault.
Symptoms: Siderail will not latch, rattles, or the CPR release is slow / stiff.
Root cause: Worn siderail latch or the CPR mechanism.
Serviceable items for this model, compiled from its documentation and engineering knowledge for this equipment class. Service class and relative cost are estimates, not manufacturer figures; no part numbers are given. Confirm intervals and pricing against the manufacturer.
Recalibrate the scale after replacement.
| Part | Service class | Relative cost | R&R difficulty |
|---|---|---|---|
| Linear actuator (backrest / knee / height) | Major service part | $$$ | Difficult |
| Central-brake / castor assembly | Major service part | $$ | Difficult |
| Load cell / scale sensor set Recalibrate the scale after replacement. | Major service part | $$$ | Difficult |
| Back-up battery | Periodic replacement | $$ | Fairly easy |
| Siderail control panel / control board | Major service part | $$$ | Moderate |
Relative cost: $ lowest · $$$$ highest service-part cost for this equipment class. Not a price.
Preventive-maintenance tasks typical of this equipment class. Perform to the manufacturer's procedure and your facility's policy - this is general reference, not a schedule for this specific unit.
A starting sequence for this model, compiled from its documentation and engineering knowledge for this equipment class. Follow the manufacturer's service documentation for the actual procedure.
Where this class of equipment is typically deployed - general reference, not this unit's own placement record.
No documents available for this device yet.
Describe a symptom or error code. Bio answers from this device's own documented error codes, failure modes, parts, and guides.
Ask Bio